85054B
3-7
Use, Maintenance, and Care of the Devices
Gaging Connectors
When to Gage Connectors
Gage a connector at the following times:
• Prior to using a device for the first time: record the pin depth measurement so that it can be
compared with future readings. (It will serve as a good troubleshooting tool when you
suspect damage may have occurred to the device.)
• If either visual inspection or electrical performance suggests that the connector interface
may be out of typical range (due to wear or damage, for example).
• If a calibration device is used by someone else or on another system or piece of equipment.
• Initially after every 100 connections, and after that as often as experience indicates.
Reading the Connector Gage
The gage dial is divided into increments of 0.0001 inch and major divisions of 0.001 inch (see
). For each revolution of the large dial, the smaller dial indicates a change of 0.01
inch. Use the small dial as the indicator of multiples of 0.01 inch. In most connector measuring
applications, this value will be zero.
When making a measurement, the gage dial indicator will travel in one of two directions. If the
center conductor is recessed from the
zero
reference plane, the indicator will move
counterclockwise to indicate the amount of
recession
, which is read as a negative value. If the
center conductor protrudes, the indicator will move clockwise to indicate the amount of
protrusion
, which is read as a positive value. Refer to
for definitions
of protrusion and recession.
Figure 3-2 Reading the Connector Gage
Содержание 85054B
Страница 1: ...Keysight Technologies 85054B 50 Type N Calibration Kit User s and Service Guide...
Страница 2: ......
Страница 3: ...85054B...
Страница 7: ...1 General Information...
Страница 15: ...2 Specifications...
Страница 20: ...2 6 85054B Specifications Electrical Specifications...
Страница 21: ...3 Use Maintenance and Care of the Devices...
Страница 29: ...85054B 3 9 Use Maintenance and Care of the Devices Gaging Connectors Figure 3 3 Gaging Male Type N Connectors...
Страница 43: ...4 Performance Verification...
Страница 46: ...4 4 85054B Performance Verification Recertification...
Страница 47: ...5 Troubleshooting...
Страница 52: ...5 6 85054B Troubleshooting Contacting Keysight...
Страница 53: ...6 Replaceable Parts...
Страница 56: ...6 4 85054B Replaceable Parts Introduction Figure 6 1 Replaceable Parts for the 85054B Calibration Kit...
Страница 57: ...85054B 6 5 Replaceable Parts Introduction Figure 6 2 More Replaceable Parts for the 85054B Calibration Kit...
Страница 58: ...6 6 85054B Replaceable Parts Introduction...
Страница 59: ...A Standard Definitions...
Страница 64: ...Index 85054B Index 4 part number 6 2 Z zeroing connector gage 3 8 3 10 3 12...