Kewtech KT1795 Скачать руководство пользователя страница 11

10.0 CLEANING AND STORAGE

   Tester does not need any special maintenance if 

used according to user manual.

   Remove tester from all test points before cleaning.
   Use a lightly damp cloth with neutral detergent 

for cleaning the instrument. Do not use abrasives 
or solvents. 

   Do not expose the instrument to direct sun light, 

high temperature and humidity or dewfall. 

   Remove batteries when the instrument will not be 

in use for a long period. 

11.  SAFETY ADVICES 

• 

 Depending on the internal impedance of the voltage 
detector there will be a different capability of 
indicating the presence or absence of operating 
voltage in case of the presence of interference 
voltage. 

• 

 A voltage detector of relatively low internal 
impedance, compared to the reference value of 
100 k�, will not indicate all interference voltages 
having an original voltage value above the ELV level. 
When in contact with the parts to be tested, the 
voltage detector may discharge temporarily the 
interference voltage to a level below the ELV, but it 
will be back to the original value when the voltage 
detector is removed. 

• 

 When the indication “voltage present” does not 
appear, it is highly recommended installing earthing 
equipment before work. 

• 

 A voltage detector of relatively high internal 
impedance, compared to the reference value of 
100 k�, may not permit to clearly indicate the 
absence of operating voltage in case of presence of 
interference voltage. 

• 

 When the indication “voltage present” appears on 
a part that is expected to be disconnected of the 
installation, it is highly recommended confirming 
by another means (e.g. use of an adequate voltage 
detector, visual check of the disconnecting point of 
the electric circuit, etc.) that there is no operating 
voltage on the part to be tested and to conclude 
that the voltage indicated by the voltage detector is 
an interference voltage.

• 

 A voltage detector declaring two values of internal 
impedance has passed a performance test of 
managing interference voltages and is (within 
technical limits) able to distinguish operating voltage 
from interference voltage and has a means to 
directly or indirectly indicate which type of voltage 
is present.

Содержание KT1795

Страница 1: ...KT1795 Current Voltage Tester Instruction Manual...

Страница 2: ...ts 7 1Voltage Test 7 2 Single pole phase test 7 3 Phase Rotation Test 7 4 Trip Test of RCD Lower ImpedanceVoltage test 7 5 Continuity test Rx Diode test 7 6 Resistance test 7 7 Torch Light 7 8 Data Ho...

Страница 3: ...instruction manual and comply with it in all sections Failure to read the tester manual or to comply with the warnings and references contained herein may result in serious bodily injury or tester dam...

Страница 4: ...voltage source such as the KEWPROVE3 opening unit Make sure that the cables tested for current are double insulated The tester must no longer be used if one or more functions fail or if no functional...

Страница 5: ...only be performed on double insulated cables If the indication voltage present appears although the checked part is considered as disconnected it is recom mended to verify additional measures if the m...

Страница 6: ...t It can be switched on with the torch light button 6 2 Auto power off Tester is automatically powered off after 30 sec when there is no signal contacted to the probes The torch light switches off aft...

Страница 7: ...ment under test aren t good enough Verification of live circuit shouldn t be dependent on this Single pole phase test only but on the voltage test Hold the tester well in your hand Connect the L2 prob...

Страница 8: ...active Continuity test automatically switches OFF after approx 30 seconds if no continuity is detected When tester is OFF If continuity is detected it will be automatically switched on again 7 6 Resis...

Страница 9: ...the battery door and replace the batteries Insert new batteries according to the engraving on the battery door Re assemble battery door Confirm that the battery door case is properly locked prior to m...

Страница 10: ...s not pressed Is 3 5mA at 1000V Peak current with push buttons pressed Is 300 mA Measurement Duty 30s ON 240s OFF Internal battery consumption approx 80mA Single pole phase test voltage range 100 1000...

Страница 11: ...o a level below the ELV but it will be back to the original value when the voltage detector is removed When the indication voltage present does not appear it is highly recommended installing earthing...

Страница 12: ...It covers all manufacturing defects for a two year period but Kewtech reserves the right to exclude abuse or accidental damage RE CALIBRATION SERVICE Regular re calibration is recommended for this ins...

Отзывы: