background image

(5) Connect the test leads to the circuit under test so the instrument is in the series.

(6) Power on the circuit under test.

(7) Measured value is indicated on the display.

6-2-4  AC Current Measurement (up to 10A)

CAUTION

● For safety sake, the measuring time on 10A range should be 15 seconds or less. Interval time between 2 measurements should be greater than 15 minutes.

If  you  carry  on  making  measurement  continuously  over  15  seconds,  or  make  measurement  in  short  interval  time,  it  may  cause  error  of  measurement  and

instrument damage.

(1) Insert the black test lead plug into the COM terminal and the red test lead plug into the 10A terminal.

(2) Set the Function Selector Switch to "A" position.(Then, "DC" , "Auto", and "A" symbols are indicated on the display.)

(3) Set the instrument to AC mode by pressing "SELECT" Key.(Then, "AC" symbol is indicated on the display.)

(4) Power off the circuit under test.

(5) Connect the test leads to the circuit under test so the instrument is in the series.

(6) Power on the circuit under test.

(7) Measured value is indicated on the display.

6-3  Resistance Measurement (Ω/ Diode check/ Continuity Check)

DANGER

● To avoid the danger of getting electrical shock, never make measurement of the circuit in which electric potential exists.

● Do not make measurement when opening the battery cover and the instrument case.

6-3-1  Resistance Measurement

(1) Insert the black test lead plug into the COM terminal and the red test lead plug into the VΩHz terminal.

(2) Set the Function Selector Switch to "Ω" position.(Then, "Auto" and "MΩ" symbols are indicated on the display.)

Make sure that the "O.L" symbol is indicated on the display at this bout, then short the test lead tips and check "0" is indicated on the display.

(3) Connect the test leads to both ends of the resistance under test.

Measured value is indicated on the display.

Note)  Even if short the test lead tips, indicated value may not be "0". But this is because of the resistance of test leads and not a failure.

In that case, by pressing "REL" key, "0" will be indicated.

6-3-2  Diode Check

(1) Insert the black test lead plug into the COM terminal the red test lead plug into the VΩHz terminal.

(2) Set the Function Selector Switch to "Ω" position.(Then, "Auto" and "MΩ" symbols are indicated on the display.)

(3) Press the "SELECT" Key twice and set the instrument to Diode Check mode.

(Then, "

" and "V" symbols are indicated on the display.)

Make sure that the ".OL" symbol is indicated on the display at this bout, then short the test lead tips and check "0" is indicated on the display.

(4) Connect the black test lead to the cathode side of the Diode and the red test lead to the anode side of the Diode.    

Forward voltage of Diode is indicated on the display.

(5) Connect the black test lead to the anode side of the Diode and the red test lead to the cathode side of the Diode.    

Normally, ".OL" symbol is indicated on the display.

Conclusion : Diode is OK if the instrument complies with above items(4) and (5).

Note)  Open-circuit voltage between measuring terminals is approx.2.8V.(measuring current approx. 0.4mA)

6-3-3  Continuity Check

(1) Insert the red test lead plug into the VΩHz terminal and the black test lead plug into COM terminal.

(2) Set the Function Selector Switch to "Ω" position.(Then, "Auto" and "MΩ" symbols are indicated on the display.)

(3) Press the "SELECT" Key one time and set the instrument to Continuity Check mode. 

― 7 ―

Содержание KT117

Страница 1: ...INSTRUCTION MANUAL AUTO RANGE DIGITAL MULTIMETER KT117...

Страница 2: ...parts in the manual for safe operation of the instrument Be sure to carefully read the instructions following each symbol in the manual DANGER is reserved for conditions and actions that are likely t...

Страница 3: ...quipment which is supplied power from a distribution board and cable run from a distribution board to an outlet 2 Features This instrument KT117 is a Digital Multimeter which designed for making measu...

Страница 4: ...600V DC or AC rms Note Forward DC current approx 0 4mA Reversed DC voltage approx 2 8V Open circuit voltage approx 0 5V Test condition Note It can not display bar graph on capacitance range It can not...

Страница 5: ...ss than 90 Insulation Resistance It should be more than 100M DC1000V between electrical circuit and enclosures Withstand Voltage It should be more than AC3700V for one minute between electrical circui...

Страница 6: ...the instrument case 6 1 1 DC Voltage Measurement DCV 1 Insert the black test lead plug into the COM terminal and the red test lead plug into the V Hz terminal 2 Set the Function Selector Switch to pos...

Страница 7: ...e circuit 5 Power on the circuit under test 6 Measured value is indicated on the display If you connect the test leads the other polarity symbol is indicated on the display 6 2 2 DC Current Measuremen...

Страница 8: ...ke sure that the O L symbol is indicated on the display at this bout then short the test lead tips and check 0 is indicated on the display 3 Connect the test leads to both ends of the resistance under...

Страница 9: ...ed on the display Measuring unit nF uF is automatically chosen and indicated due to the measured value Note It may take some time according to the measuring capacitance Measuring capacitance 4uF Measu...

Страница 10: ...uto ranging there are following 2ways 1 Press RANGE Key for 2seconds 2 Change to the other functions 7 4 REL Key The difference between measured values can be indicated on the display at each ACV DCV...

Страница 11: ...e to remove test leads from the instrument when opening the battery cover in order to replace batteries and fuses 9 1 Battery replacement 1 Remove test leads from the instrument 2 Remove the holster f...

Страница 12: ...92 1819A...

Страница 13: ...Impedance 10M 1 5 5dgt 6 000V 1 2 3dgt 60 00 600 0V 1 5 5dgt 600V DC A 600 6000 A 60 600mA 6 10A 1 2 3dgt 600 6000 A 60 600mA 2 0 5dgt 6 10A AC A 600 6000 A 60 600mA 6 10A 1 5 4dgt 600 6000 A 60 600mA...

Отзывы: