background image

Contents

1 Functions and accuracy ......................................................................................... 2
2 What’s in the box?  ................................................................................................. 3
3 Measuring reference values for ongoing accuracy records  ...................................... 3
4 Checking Continuity performance  .......................................................................... 4
 

4.1 The need to null test leads  ........................................................................ 4

 

4.2 Instruments with a lead null function  .......................................................... 4

 

 

4.2.1 Lead nulling  ............................................................................. 4

 

 

4.2.2 Continuity check  ....................................................................... 4

 

4.3 Instruments without a lead null function  ...................................................... 5

 

 

4.3.1 Measuring lead resistance  ......................................................... 5

 

 

4.3.2 Continuity check  ....................................................................... 5

 

4.4 Tolerance examples  ................................................................................. 5

5 Insulation Resistance checks  ................................................................................. 6
6 Earth Fault Loop Impedance checks  ...................................................................... 7
7 Residual Current Device checks  ............................................................................. 9

1 Functions & accuracy

The Kewtech FC2000 provides a quick and simple method of checking the accuracy of 
electrical installation test instruments.
It does not replace the calibration requirements but provides a ready check on instrument 
accuracy helping to minimise incorrect certification and costly revisits to confirm installation 
compliance.
The FC2000 provides accuracy checks for test instruments used to measure:
• Continuity
• Insulation Resistance
• Earth fault loop impedance
• Operation of Residual Current Devices.

Accuracy

2   What’s in the box?

1 × FC2000
1 × Power Lead
2 × Test leads for Continuity and Insulation Resistance checks
Calibration Certificate
The FC2000 is supplied with a Calibration Certificate. The measured values given on the 
Check Label fitted in the lid and shown on the Calibration certificate are the actual resistance 
values for each individual FC2000. These values may vary from the Nominal Value due to the 
manufacturer’s tolerances for the resistors used.

3   Measuring reference values for ongoing accuracy  

records

The FC2000 is supplied with a Calibration Certificate. The measured values given on the 
Check Label fitted in the lid are the ‘actual’ resistance values for each individual FC2000. 
These values may vary from the Nominal Value due to the manufacturer’s tolerances for the 
resistors used.
When using the FC2000 for the first time each test instrument to be checked should be in 
current calibration.
When the first set of tests for any instrument are undertaken:
•   A separate test record sheet should be kept for each test instrument.
•   Each test should be repeated 3 times to establish the initial test value for each particular  
  instrument.
•   The value obtained over the three tests should be similar and within the instrument  
  manufacturer’s tolerances.
•   The average value over the three tests should be recorded on the instrument test record 
  as the Reference Value.
•   It is this Reference Value which will be used for future accuracy checks on the instrument.

Insulation Test

0.5MΩ

1 %

Voltage Rating to 

1000V.

2.0MΩ

1%

200MΩ

2%

Continuity Test

0.5Ω

1% ± 20mΩ

Current Rating

1.0Ω
2.0Ω

Loop Test

additional 1Ω or 

100Ω switched in to 

the Earth Loop

1% ± 20mΩ

Current Rating (pulsed)

RCD

x 1

Current 1% of 

range

Timer ± 1ms

x 5

Check Value

Acceptable range

Rating

30mA x 1 (30mA)

30 to 33mA

30mA AC type

30mA x 5 (150mA)

143 to 165mA

30mA AC type

30mA x 1/2 (15mA)

13.5 to 15mA

30mA AC type