(4)   When the start/stop switch at the upper left of the main unit or 

the

 key is pressed, the next measurement may be made.

* Note 1:   Even if the printer is still printing, as long as the motor has stopped, 

the next measurement may start. 

* Note 2:   To stop a test in the middle of a measurement, please press the 

"start/stop" switch. The average moisture value for the number of 
kernels measured up to that point will be displayed. 

* Note 3:   When kernels have not been counted within 30 or more seconds, 

the buzzer will sound and "POURING" will be displayed. Since the 
sample may have run out, or the sample may have plugged the 
screw feed, please check the interior of the drum. This display will 
disappear after sample is again loaded and the count starts.

Содержание PORQUE-520

Страница 1: ...Single Kernel Grain Moisture Tester PQ 520 type PQ 5205 1 Operating Instructions...

Страница 2: ...age Excessive voltage can cause damage or fire Be sure to ground the electrical power cord When the electrical outlet is a two prong type please connect the green colored ground line protruding from t...

Страница 3: ...4 4 Setting the Number of Kernels to Measure 7 5 Measurement 8 5 1 Loading the Sample 8 5 2 Starting Measurement 8 6 Measurement Precautions 10 6 1 Cleaning the Internal Feed Unit 10 6 2 Temperature C...

Страница 4: ...aking it possible to accurately determine the moisture distribution of the sample and preventing uneven moisture caused by inadequate drying Features 1 Simple operation 2 Measurements may be quickly p...

Страница 5: ...f average moisture value Measurement Precision 0 5 moisture less than or equal to 20 Display Method Fluorescent tube display Contents of Display Selected grain type average moisture value kernel count...

Страница 6: ...eration Panel Discharge Tray Power Supply Switch 3 Names of Parts Fan USB Port Printer Connection Connector Electrical Power Inlet Fuseholder Printer Connection Electrical Power Supply Outlet AC 100V...

Страница 7: ...be measured Set number of kernels Use histogram to display the moisture distribution Start or stop measurement Display Contents Grain Type Average Moisture Value Number of Kernels Time Histogram Feed...

Страница 8: ...other countries Please only use a power cord designed for your country 1 Insert the power cord into the power supply connector on the rear side of the main unit 2 If the power supply is a three prong...

Страница 9: ...kernels Note 1 This is set to 100 kernels at the time of shipment of the unit Note 2 If the number is set to less than 10 kernels the setting will become 10 kernels Note 3 The number of kernels that c...

Страница 10: ...cated on the right Note 1 Measurement is not possible without proper placement of the main unit internal transparent panel If PANEL is displayed please check the transparent panel Note 2 The motor sto...

Страница 11: ...of a measurement please press the start stop switch The average moisture value for the number of kernels measured up to that point will be displayed Note 3 When kernels have not been counted within 3...

Страница 12: ...ed when the motor is rotating rotation will stop 3 Lift the lug of the internal transparent panel and pull the internal panel out to the front Note Breakage may result if the transparent panel is forc...

Страница 13: ...of the instrument Temperature compensation is automatically performed However when the difference between the temperature of the grain and the temperature of the machine is too large temperature comp...

Страница 14: ...ndard deviation temperature 2 yyyy mm dd type of grain no of measured kernels average moisture standard deviation temperature histogram 3 yyyy mm dd type of grain no of measured kernels average moistu...

Страница 15: ...13 7 3 Sample Print out Print Mode 1 Print Mode 2 Print Mode 3 Kernel count scale of the histogram Kernel count scale of the histogram Total Moisture Data...

Страница 16: ...rocedure enter the month day hour minute 7 After entering the minute last position the time is confirmed by pressing the key The unit returns to the pre measurement state Note The second count becomes...

Страница 17: ...sor temperature is greater than or equal to 70 C The thermistor is malfunctioning Repair is needed There is a problem with the electrical circuit Repair is needed PANEL The transparent panel isn t att...

Страница 18: ...nals The moisture value is high or low There is the possibility that the measured sample has not reached the temperature of the unit Please wait until the sample has reached room temperature The time...

Страница 19: ...eriodic Inspection It is recommended that this instrument undergo periodic inspection in order to check and maintain performance of this product Although timing will depend on the frequency of use of...

Страница 20: ...accessory items mentioned in this document may differ from those of the actual products and accessory items although operation and function will not be affected Although the contents of this document...

Отзывы: