![Kett LE-373 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/kett/le-373/le-373_operating-manual_1958603014.webp)
14
1
Application
LE-373 & LH-373 : A total of 100 applications (calibration
curves) can be set.
LZ-373 : A total of 100 applications , consisting of
application numbers 0 to 49 for the electromagnetic
type measurement and application numbers 50 to 99
for the eddy-current type measurement, can be set.
・・
・・
(1) Perform steps (1) and (2) on p.13.
(2) Press the Del key, and after the
displayed application number
bar has been deleted, enter the
desired application number.
Then press the Enter key to
apply your selection and return
to the measurement screen.
* Once an application number has been
set, it is saved even after the power
is switched off until the next time it is
changed.
* LZ-373 : The selectable application
number range is 0 to 49 for Fe and 50 to
99 for NFe. If an out-of-range number is
selected, an error is displayed.
(3) To leave the application number
unchanged, select
Esc
with the
→
key, and then press the Enter
key. The screen will change
b a c k t o t h e m e a s u r e m e n t
screen without changing the
application number.
2
Substrate Cal.
Once substrate calibration has been done, it is not
necessary to perform calibration for subsequent
measurements. However, if the object to be measured
is changed or the probe is replaced, etc., calibration
must be performed again.
* A substrate refers to a piece made with the same material
and shape as the pieces to be measured, but not covered
with coating (plating or paint).
(1) Perform steps (1) and (2) on p.13.
(2) Point the probe in the air and
press the Enter key.
Measurement is per formed
automatically 7 times.
Содержание LE-373