5-1.  Connecting the Power Cord

[Caution]  Power supply voltage may be different when 

using this product in other countries. Please only  

use a power cord designed for your country.

(1)  Insert the power cord into the power supply connector on 

the rear side of the main unit.

(2)  If the power supply is a three-prong plug 100V-120V outlet, 

insert the power cord into VAC outlet. (See the right fig and 

in case of 100V-120V uses with blade plug condition)

 

Connect the included power plug conversion adapter into 

the power cord first, then plug the cord into a 100V-120V 

power outlet. Next, connect the ground wire on the power 

cord to the ground.

[Caution]   200V-240V outlet is used, power cord B should be used.

[Caution]  If the power supply is 200V-240V with flat blade plug, the 

conversion adapter should be prepared by yourself.

[Caution]   If the power supply is 100V-120V with round pin plug, the 

conversion adapter should be prepared by yourself.

5-2.  Inserting the Whiteness Standard Plate

 

As shown in the figure, insert the whiteness standard plate 

into the sample case slot on the main unit until it cannot be 

inserted any further.

[Caution]  The whiteness standard plate lid should never be 

opened unless you need to clean the glass surface.

[Caution]  Make sure the rice whiteness plate is facing in the 

correct direction.

5-3.  Starting Up the Main Unit

 

Turn the power switch located on the rear side of the main 

unit to ON. "WARMING-UP" is displayed. You will see 

numbers flashing and after approximately 20 seconds, the 

value of the whiteness standard plate is displayed. Once 

the value of the whiteness standard plate is displayed, 

remove the whiteness standard plate.

[Caution]  If the whiteness standard plate is not inserted correctly 

when the power switch is turned ON, the display will 

repeatedly show "INSERT STD. "If this should occur, 

insert the whiteness standard plate into the sample case 

slot on the main unit until it cannot be inserted any further. 

 

The C-600 is now ready for measurement.

5. Preparing for Measurement

Sample Case Slot

Whiteness Standard Plate

R

e

a

r Sid

e

Содержание C-600

Страница 1: ...Rice Whiteness Tester C 600 User Manual...

Страница 2: ...read the precautions included in the user manual Please don t use this product if it is not working properly If the product should break or malfunction please contact the retail outlet where you purch...

Страница 3: ...Principles and Characteristics 4 2 Specifications 5 3 Part Names 6 4 Description of Operation Keys 7 5 Preparing for Measurement 8 6 Measurement 9 7 Maintenance and Other Functions 11 8 Errors 14 Tabl...

Страница 4: ...light reflected back and detected and thus a higher measurement value is registered The whiteness of rice correlates directly with the milling percentage of that rice Therefore you can use this devic...

Страница 5: ...30 to 85 non condensing Light Source Blue LED External Output RS 232C Power Source AC 100V 120V 50 60Hz Power cord A Flat blade attachment plug Type A 1 AC 200V 240V 50 60Hz Power cord B Round pin att...

Страница 6: ...03 C 600 04 C 600 05 Main Unit Cover Glass Wipe Replacement Sponge Replacement Packing x3 Replacement Fuse C 600 06 C 600 07 C 600 08 C 600 09 C 600 10 Power Cord A Power Cord B Power Plug Conversion...

Страница 7: ...r output ON or OFF Used as a backspace when inputting numerical values Used to input standard plate values Used to finalize numerical input Used to input numerical values Used for sensitivity calibrat...

Страница 8: ...f 5 2 Inserting the Whiteness Standard Plate As shown in the figure insert the whiteness standard plate into the sample case slot on the main unit until it cannot be inserted any further Caution The w...

Страница 9: ...the rice from the cup through the hole in the fixed quantity shooter Insert all of the rice from the cup until the rice begins to overflow out of the hole in the fixed quantity shooter Caution The sam...

Страница 10: ...ents is displayed after each measurement is completed However the number of measurements is reset to one in the following cases 1 When the key is pressed 2 When the number of measurements exceeds nine...

Страница 11: ...he sample platter sponge wears the amount of sample material you can fit into the sample platter increases which results in inaccurate measurements The sponge needs to be replaced when you can see lin...

Страница 12: ...s played remove the whiteness standard plate Note This procedure must be performed whenever the whiteness standard plate is replaced 7 5 Replacing the Fuse 1 Turn the power OFF and disconnect the powe...

Страница 13: ...NO PRINT is displayed printer output is turned OFF Caution Connect the device to the printer using the printer cable VZC 14 only after setting up the printer VZ 330 Caution Refer to the printer s use...

Страница 14: ...ion Insert the whiteness standard plate then try performing the sensitivity calibration again Circuitry error Repair is required Circuitry error Repair is required Circuitry error Repair is required C...

Страница 15: ......

Страница 16: ...nd parts depicted in this user manual may not appear exactly as their actual counterparts but this does not affect their operation or functionality This user manual was intended to be written as clear...

Отзывы: