background image

11

7-1.  Main Unit Glass Filter

 

If the glass filter on the main unit is dirty, you will not receive 

an accurate measurement. You must keep the glass filter 

clean at all times when using the device. Use your finger  

to lift upwards on the handle on the inside of the front cover,  

in order to remove the front cover. This gives you access  

to the glass filter for cleaning.

[Caution]  There are sharp edges protruding from the front 

cover and the metal parts of the main unit. Be care-

ful not to injure yourself when cleaning the glass 

filter.

7-2.  Sample Platter Sponge

 

As the sample platter sponge wears, the amount of sample 

material you can fit into the sample platter increases, which 

results in inaccurate measurements. The sponge needs  

to be replaced when you can see lines on the inside of the 

sample platter when it is empty. The sponge is attached  

to the bottom of the platter with double-sided tape. Simply 

pull out the sponge to replace it.

7-3.  Main Unit Glass Filter Packing

 

When the packing attached to the main unit's glass filter 

wears down, please replace it. If this packing wears down,  

a space will form between the glass filter and the sample case.

[Caution]  Attach the packing in the orientation shown in the fig

-

ure.

7. Maintenance and Other Functions

Sponge

Worn Packing

Narrow

Wide

Caution

Front Cover

Handle

Glass Filter

Содержание C-600

Страница 1: ...Rice Whiteness Tester C 600 User Manual...

Страница 2: ...read the precautions included in the user manual Please don t use this product if it is not working properly If the product should break or malfunction please contact the retail outlet where you purch...

Страница 3: ...Principles and Characteristics 4 2 Specifications 5 3 Part Names 6 4 Description of Operation Keys 7 5 Preparing for Measurement 8 6 Measurement 9 7 Maintenance and Other Functions 11 8 Errors 14 Tabl...

Страница 4: ...light reflected back and detected and thus a higher measurement value is registered The whiteness of rice correlates directly with the milling percentage of that rice Therefore you can use this devic...

Страница 5: ...30 to 85 non condensing Light Source Blue LED External Output RS 232C Power Source AC 100V 120V 50 60Hz Power cord A Flat blade attachment plug Type A 1 AC 200V 240V 50 60Hz Power cord B Round pin att...

Страница 6: ...03 C 600 04 C 600 05 Main Unit Cover Glass Wipe Replacement Sponge Replacement Packing x3 Replacement Fuse C 600 06 C 600 07 C 600 08 C 600 09 C 600 10 Power Cord A Power Cord B Power Plug Conversion...

Страница 7: ...r output ON or OFF Used as a backspace when inputting numerical values Used to input standard plate values Used to finalize numerical input Used to input numerical values Used for sensitivity calibrat...

Страница 8: ...f 5 2 Inserting the Whiteness Standard Plate As shown in the figure insert the whiteness standard plate into the sample case slot on the main unit until it cannot be inserted any further Caution The w...

Страница 9: ...the rice from the cup through the hole in the fixed quantity shooter Insert all of the rice from the cup until the rice begins to overflow out of the hole in the fixed quantity shooter Caution The sam...

Страница 10: ...ents is displayed after each measurement is completed However the number of measurements is reset to one in the following cases 1 When the key is pressed 2 When the number of measurements exceeds nine...

Страница 11: ...he sample platter sponge wears the amount of sample material you can fit into the sample platter increases which results in inaccurate measurements The sponge needs to be replaced when you can see lin...

Страница 12: ...s played remove the whiteness standard plate Note This procedure must be performed whenever the whiteness standard plate is replaced 7 5 Replacing the Fuse 1 Turn the power OFF and disconnect the powe...

Страница 13: ...NO PRINT is displayed printer output is turned OFF Caution Connect the device to the printer using the printer cable VZC 14 only after setting up the printer VZ 330 Caution Refer to the printer s use...

Страница 14: ...ion Insert the whiteness standard plate then try performing the sensitivity calibration again Circuitry error Repair is required Circuitry error Repair is required Circuitry error Repair is required C...

Страница 15: ......

Страница 16: ...nd parts depicted in this user manual may not appear exactly as their actual counterparts but this does not affect their operation or functionality This user manual was intended to be written as clear...

Отзывы: