Kett AN-920 Скачать руководство пользователя страница 17

17

<Reference 1>

● Display of protein's constant moisture basis (CM)

The protein value to be displayed with this unit indicates the percentage of 

constituent over a whole measuring object including moisture. It is useful, 

for example, when calculating the weight of protein contained in the meas-

ured grain. However, when comparing the protein value in the multiple 

targets, there may be a case it is more convenient to assume that the 

targets have the fixed moisture value. The CM display of the protein value is 

based on this concept. The CM indicates the converted protein value 

assuming that the moisture value of the measurement targets is a certain 

given value ("@Moisture%" with AN-920). When the given moisture value is 

"0%", it is called "Dry Moisture basis" (DM). The conversion formula is as 

follows. To display correct "Protein CM", the calibration curves of "Protein" 

and "Moisture" and "@Moisture%" needs to be set properly. 

                          Protein × (100-@Moisture%)

Protein CM = 

 

                    100 - Moisture

<Appendix> To use the unit in the best accuracy

Regularly check the standard samples in order to minimize the affects to the measurement by the 

changes that may occur in the time course. In addition, the standard sample should be checked 

when the measurement environment is changed, the operator is changed, after a power failure, and 

when the measured value seems not reliable.

■ Check using standard samples

● What is standard sample?

A standard sample is the sample whose standard values have been calibrated before shipment.

Each standard value is indicated on the top of the standard sample case.

● Storage

•   Store the sample in the cool dark place (5 to 10°C) such as in the refrigerator.

•   Never store the sample in the freezer. The sample will be frozen.

•   Other than the appropriate storage, it may cause a change to the standard value.

•   The validation period indicated on the standard sample is applicable as long as the sample is 

appropriately stored.  

Use only the sample within the validation period.

● How to check using standard samples

(1) Take out the sample from the cool dark place before a half day before the measurement.

(2)  Place the AN-920 main unit and the standard sample in a same place to have them thermally 

stabilized.

(3) Measure the standard sample for three to five times, and calculate the average value.

(4) Compare the measured composition value with the standard value of each composition.

Example: 

Reference

Average of measurement values

Difference

Standard 

sample

16.0%

16.2%

0.2%

 

 

 In this example, the difference is "0.2". Therefore, a bias adjustment is not necessary in the 

standard use. If the difference is over "0.2", repeat the above procedure and compare the 

values again. As a result, if the difference is over "0.2", a bias adjustment is required. Each 

composition has different limit value. We recommend the following an a reference. Define 

the limit value according to the operating environment.

<Reference 2> 

●  Display of Quality Evaluation Value 

(Q.E.V.)

While "Protein CM" is not displayed, the Q.E.V. 

will be calculated from the measured protein 

value (not CM) and the moisture included value. 

When "@Moisture%" is set less than 10% and 

"Protein CM" is displayed, the Q.E.V. will be 

calculated from the measured protein value (not 

CM) and the moisture included value. When "@

Moisture%" is set more than 10% and "Protein 

CM" is displayed, the Q.E.V. will be calculated 

replacing "Protein" with "Protein CM" and "Mois-

ture" with "@Moisture%".

Composition

Reference limit value

Protein

0.3%

Moisture content

0.2%

Amylose (ref.)

1.0%

Содержание AN-920

Страница 1: ...NIR Rice Composition Analyzer AN 920 Operating Manual Thank you for purchasing this product Please read the operating manual carefully and use this product properly...

Страница 2: ......

Страница 3: ...ts of Main Unit 6 2 3 Accessories 7 3 SPECIFICATIONS 8 4 DESCRIPTION OF OPERATION KEYS 9 5 EASY OPERATION GUIDE 10 6 ADJUSTMENT 12 7 SETUP 13 8 OTHERS 16 8 1 Printer Output 16 8 2 Replacing Fuses 16 8...

Страница 4: ...a place where explosive inflammable gas or liquid substance exists Otherwise it may cause explosion fire or electric shock etc Do not use other than rated power supply voltage If overvoltage is applie...

Страница 5: ...le or other cables never pull on the cord or cable and instead hold the plug or cable connector Do not install the unit in a place where handling the power cable and connectors or turning the power sw...

Страница 6: ...e a tungsten lamp is used High measurement accuracy can be achieved by adopting a method in which the transmitted light from the sample is detected after being diffracted into 600 to 1100 nm by a spec...

Страница 7: ...ccessories Spare fuse Sampling cup Power cable Sample case 2 Operating manual Storage container Standard sample Storage container Standard sample brown rice milled rice 1 each Stored in the storage co...

Страница 8: ...ns and weight 230 W 400 D 250 H mm 8 5 kg Accessories Power cable Fuse Sample case 2 Sampling cup Standard sample brown rice milled rice 1 each storage container standard sample operating manual Optio...

Страница 9: ...he value 7 SETUP P 13 14 15 Display date time communica tion light intensity reset and user switching can be set 6 ADJUSTMENT P 12 Adjustments are required before starting the measurement There are tw...

Страница 10: ...ommended to keep the tempera ture of the sample and the environ mental temperature close Clean the sample case If the window of the sample case is not clean or significantly scratched adjustment or me...

Страница 11: ...ver open the cover 15 The measurement result will be displayed Pressing the MEASURE key in this screen will enable a new measurement Pressing the Home key will restore the initial screen The average o...

Страница 12: ...anually calculated Change the value with keys and then press the Enter key 4 Pressing the Home key will restore the initial screen Procedure 1 Press the Adjustment key 2 The measured value of the stan...

Страница 13: ...he Setup key 2 Press the DISPLAY key 3 For each item press to display and to hide 4 To set the display range press the value Change the value with keys and press the Enter key When Protein CM is set t...

Страница 14: ...When the tungsten lamp is blown replace the lamp in the following procedure Make sure to purchase the new lamp from Kett Do not use other lamp than specified one While the power is turned on the lamp...

Страница 15: ...asurement Calibration curve selection Bias adjustment Display setting Date Time setting Communica tion setting Light intensity resetting User addition Administrator User Guest 3 Switching the user Pro...

Страница 16: ...easurement Sample is too dark bright Instrument temperature is too low high Sample temperature is too low high A D Converter Error Replace lamp The following messages may be displayed during initializ...

Страница 17: ...refrigerator Never store the sample in the freezer The sample will be frozen Other than the appropriate storage it may cause a change to the standard value The validation period indicated on the stand...

Страница 18: ......

Страница 19: ......

Страница 20: ...and parts depicted in this user manual may not appear exactly as their actual counterparts but this does not affect their operation or functionality This user manual was intended to be written as clea...

Отзывы: