2.   After removing the used battery, insert a new battery in the 

correct direction of the battery terminal.

3.   Place the battery cover on the back of the device, screw it in, 

and insert the silicone ball into the screw hole.

PRECAUTIONS

•  Dispose of used batteries in accordance with applicable laws 

and regulations in the current country and region.

•  Contact the dealer, Kerr Endodontics sales representative or Kerr 

Customer Care and use an Apex Connect Battery approved by 

Kerr Corporation. 

CLEANING, DISINFECTION AND STERILIZATION 

INSTRUCTIONS

After using the device, be sure to keep it thoroughly clean for the 

next use. And clean everything thoroughly before autoclaving. In 

order to prevent the dirt on the surface of the device from drying 

out before cleaning, please wash it as soon as possible after use. 

In addition, the reprocessing procedure should minimize delays 

between steps.

Autoclavable Components: Lip hook, File 
holder A and File holder B

The lip hook and the file holder must be sterilized after cleaning.

 

1

  Cleaning and Disinfection: File Holder A

1.   Put the file holder A in the tap water for 5 minutes.
2.   Thoroughly clean the surfaces of the file holder A with a 

tissue containing isopropyl alcohol (e.g., CaviWipes).

   

Wipe the surface of the file holder A at least twice in  

one minute.

   

Clean the surface of the file holder A thoroughly to avoid 

leaving any residue of contamination. 

3.   Rinse the file holder A in tap water for at least 30 seconds.
4.   Dry the file holder A with a soft cloth, and dry at room 

temperature for at least 30 minutes.

 

2

  Cleaning and Disinfection: File Holder B

NOTE: Do not spray the File Holder B with a liquid 

disinfectant.

1.   Clean the surfaces of the file holder B with a tissue 

containing isopropyl alcohol (e.g., CaviWipes).

•  Wipe the surface of the file holder B at least twice in  

one minute.

•  Clean the surface of the file holder B thoroughly to avoid 

leaving any residue of contamination.

2.   Dry the file holder B with a soft cloth, and dry at room 

temperature for at least 30 minutes.

 

3

  Cleaning and Disinfection: Lip Hook

1.   Put the lip hook in the tap water for 5 minutes.
2.   Thoroughly clean the surfaces of the lip hook with a tissue 

containing isopropyl alcohol (e.g., Caviwipes).

•  Wipe the surface of the lip hook at least twice in  

one minute.

•  Clean the surface of the lip hook thoroughly to avoid 

leaving any residue of contamination.

3.   Rinse the lip hook in tap water for 30 seconds.
4.   Dry the lip hook with a soft cloth, and dry at room 

temperature for at least 30 minutes.

 

4

  Packing and Sterilization

NOTE:

   

Health Care facilities are responsible for making sure that 

the sterilization equipment is calibrated according to the 

manufacturer’s manuals and specifications. In addition, 

health care facilities are responsible for training their staff 

on infection control, proper sterilization and disinfection 

procedures

   

Make sure that the sterilization pouches are suitable for 

steam sterilization and comply with the national guidelines, 

standards, and requirements.

 

ISO 11607

 

For USA: Use FDA-cleared accessories

1.   Place the cleaned lip hook and the file holder(s) into an 

autoclave pouch.

2.   Recommended sterilization parameters:

Cycle

Gravity 

Displacement

Pre-Vacuum

Temperature (°C)

121°C (250°F)

132°C (270°F)

Exposure time (min)

30 minutes

4 minutes

Drying time (min)

30 minutes

30 minutes

3.   Take out the lip hook and the file holder from the autoclave 

and dry it while keeping it packed in the pouch for at least 

30 minutes at room temperature.

PRECAUTIONS

•  Maximum number of re-sterilization of lip hook and the file 

holder is 250 times.

•  Immediately after sterilization, the lip hook and the file holder 

can be very hot. Cool it sufficiently before use as there is a risk of 

injury such as burns.

•  Do not leave the lip hook and the file holder inside the 

autoclave after sterilization. 

•  Never sterilize components other than the lip hook and the 

file holder.

NON-AUTOCLAVABLE COMPONENTS

 

1

  Cleaning and Disinfection: Apex Connect Unit

After using the device, be sure to keep it clean for the next use.

   

Wipe the surface of the Apex Connect Unit using a cloth  

(or gauze) lightly moistened with ethanol (70-80%).  

Clean the crevices with a cotton swab moistened with 

ethanol. And dry at room temperature for 30 minutes.

PRECAUTIONS

•  Do not use organic solvents such as thinner, benzene, or 

methanol to clean the device. 

•  Do not put the device in alcohol or water and be careful not to 

let foreign substances such as water or dust get inside the device.

 

2

  Cleaning and Disinfection: Probe Cord

1.   Wet a small soft brush in ethanol for disinfection (ethanol  

70-80 vol%) to remove all contaminants from the crevices  

and all surfaces. (repeat 20 times)

2.   Wipe all surface thoroughly using a cloth (or gauze) lightly 

dampened in ethanol for disinfection. (repeat 20 times)

3.   Wipe all surfaces with a cloth moistened with distilled 

water, then dry with a clean, dry cloth. 

4.   After visual inspection to ensure cleanliness, repeat the 

cleaning steps if contamination is visible.

Содержание Apex Connect

Страница 1: ...Apex Connect INSTRUCTIONS FOR USE ELECTRONIC APEX LOCATOR...

Страница 2: ...buttontoenterthesettingsmode Inthesettingsmode theAPEXcanbesetforthereferencevalue soundvolume andstandbytime 5 Selectbutton Inthesettingsmode changethesettings 6 Batterycover Thereisabuilt inrecharg...

Страница 3: ...cause a malfunction Ensure conductors such as metal are not in contact with the metal part of the device There is a risk of device malfunction fire or electric shock Do not disassemble the device Dire...

Страница 4: ...he display When charging is complete the battery icon stops blinking and stays on WARNING ChargeusingtheAC DCadapterandpowercordincludedin thepackage Besuretoconnectthepowerplugtoagroundedpowersupply...

Страница 5: ...idity 30 75 Atmosphere pressure 700 1060 hPa Do not use the device in a place that it is exposed to direct sunlight heat sources or near a fire 2 Assembly Charge the battery sufficiently when using th...

Страница 6: ...ed from the measuring device may be shown This is not a malfunction of the measuring device It may be due to the anatomical apex and or the physiologic apex may be located coronal to the radiographic...

Страница 7: ...NOTE For more information on the components of the Elements Connect Cordless Endodontic Motor and how to use them refer to the Elements Connect Endodontic Motor Instructions for use 3 Assembly and Co...

Страница 8: ...sasfollows R S Themotorstopsafterrotatingintheopposite directionofthecurrentsetting REV Themotorrotatesinthedirectionoppositetothe currentsetting Itthenrotatesinthesetdirectionagain 2 Itisthetorqueval...

Страница 9: ...e of contamination 3 Rinse the lip hook in tap water for 30 seconds 4 Dry the lip hook with a soft cloth and dry at room temperature for at least 30 minutes 4 Packing and Sterilization NOTE Health Car...

Страница 10: ...d errorsmayoccur so pleasestartthemeasurement withathickfiletoathinfile Thisisamulti rootedcanal tooth Checkwitharadiographifthere areanycracksintheteethand thenre measure Problem Cause Solution Erwil...

Страница 11: ...15 ApexConnectFileHolderA 815 1816 ApexConnectFileHolderB 815 1817 ApexConnectReplacementBattery ELECTROMAGNETIC COMPATIBILITY This device has been tested for compliance with electromagnetic tests acc...

Страница 12: ...errdental com Manufactured by Meta Systems Co Ltd 1214 18 Sicox tower 12F 484 Dunchon daero Jungwon gu Seongnam si Gyeonggi do 13229 Korea Distribution Facility Kerr Australia Pty Ltd 12 Mars Rd Unit...

Отзывы: