KERR ENDODONTICS Apex Connect Скачать руководство пользователя страница 7

5

6

•  APEX reference value can be set between -0.5 and 1.0 mm. 

(Default: 0.0 mm)

•  The default setting of the APEX reference value which is 0.0 

means that the file has reached the anatomical foramen 

(apical constriction) and is approximately 0.5 mm away 

from the physiological foramen (apical foramen).

3.   Press the 

[ ]

 button to select the alarm volume, and then 

press the 

[ ]

 button. The settings are saved, and the  

value flashes.

•  Set the alarm volume between 0 and 3. (Default: 2)

4.   Press the 

[ ]

 button to select the standby time, and then 

press the 

[ ]

 button. The settings are saved and returned to 

the measurement screen.

•  Set a standby time so that the device turns off after a 

certain period of inactivity.

•  The standby time can be set between 2 and 10 minutes, in 

2-minute increments. (Default: 4 minutes)

ADVANCED SETTINGS

 

1

   Linking the Apex Connect to the Elements 

Connect Endodontic Motor 

This device can be used in conjunction with Elements Connect 

Endodontic Motor, by connecting the two devices, the 

instrumentation and enlargement of the root canal can be 

performed while checking the working length of the root canal.

PRECAUTIONS

•  This device is designed to be compatible with Elements Connect. 

Do not connect any device other than Elements Connect.

•  Use the USB connection terminal on this device only to connect 

Elements Connect and lip hook. Do not connect the charging 

USB cable.

 

2

   Elements Connect Endodontic Motor 

Components

The following components are required to set up the Cordless 

endodontic motor:

Contra angle

Motor Handpiece

Battery (2)

NOTE: 

For more information on the components of the Elements 

Connect Cordless Endodontic Motor and how to use them, refer 

to the 

Elements Connect Endodontic Motor Instructions 

for use.

 

3

  Assembly and Connection

1.   Plug the USB connector (H) on one end of the probe cord into 

the USB connection terminal of the motor handpiece and the 

other USB connector (A) into that of the Apex Connect Unit.

PRECAUTIONS

The USB connector (Type C) has specific connectors for each 

intended use. Examine the role of each connector and connect to 

the appropriate device.
•  Single USB connector (A): Apex Connect device connection. 
•  USB connector (H) supplied with lip hook connector: Motor  

handpiece connection. 

2.   Insert the lip hook into the lip hook connector of the probe cord.

PRECAUTION

Be sure to clean and sterilize the lip hook, File Holder A and File 

Holder B before use. Refer to 

Cleaning, Disinfection and 

Sterilization Instructions

3.   Refer to 

BEFORE USE > Using Accessory

 in the Elements 

Connect Cordless Endodontic Motor Instructions for use 

to attach the contra angle and Ni-Ti file to the motor 

handpiece. 

PRECAUTION

When selecting the proper file length for the procedure, note that 

the total working length of the file is reduced by 2mm due to the 

deep set insert in the contra angle head and might difficult the 

repositioning of the silicone stopper. A longer file might have to 

be selected. 

4.   Press the 

[Main]

 button on the motor handpiece and the  

[ ]

 button on the Apex Connect Device. The power and the 

display of both turn on.

5.   Check that 

[EAL]

 appears on the main screen of the motor 

handpiece. If 

[EAL]

 appears, it means that the motor 

handpiece and main body are properly connected.

 

4

  Setting the Linkage

Set the motor operation options when using the device in 

conjunction with Apex Connect

Содержание Apex Connect

Страница 1: ...Apex Connect INSTRUCTIONS FOR USE ELECTRONIC APEX LOCATOR...

Страница 2: ...buttontoenterthesettingsmode Inthesettingsmode theAPEXcanbesetforthereferencevalue soundvolume andstandbytime 5 Selectbutton Inthesettingsmode changethesettings 6 Batterycover Thereisabuilt inrecharg...

Страница 3: ...cause a malfunction Ensure conductors such as metal are not in contact with the metal part of the device There is a risk of device malfunction fire or electric shock Do not disassemble the device Dire...

Страница 4: ...he display When charging is complete the battery icon stops blinking and stays on WARNING ChargeusingtheAC DCadapterandpowercordincludedin thepackage Besuretoconnectthepowerplugtoagroundedpowersupply...

Страница 5: ...idity 30 75 Atmosphere pressure 700 1060 hPa Do not use the device in a place that it is exposed to direct sunlight heat sources or near a fire 2 Assembly Charge the battery sufficiently when using th...

Страница 6: ...ed from the measuring device may be shown This is not a malfunction of the measuring device It may be due to the anatomical apex and or the physiologic apex may be located coronal to the radiographic...

Страница 7: ...NOTE For more information on the components of the Elements Connect Cordless Endodontic Motor and how to use them refer to the Elements Connect Endodontic Motor Instructions for use 3 Assembly and Co...

Страница 8: ...sasfollows R S Themotorstopsafterrotatingintheopposite directionofthecurrentsetting REV Themotorrotatesinthedirectionoppositetothe currentsetting Itthenrotatesinthesetdirectionagain 2 Itisthetorqueval...

Страница 9: ...e of contamination 3 Rinse the lip hook in tap water for 30 seconds 4 Dry the lip hook with a soft cloth and dry at room temperature for at least 30 minutes 4 Packing and Sterilization NOTE Health Car...

Страница 10: ...d errorsmayoccur so pleasestartthemeasurement withathickfiletoathinfile Thisisamulti rootedcanal tooth Checkwitharadiographifthere areanycracksintheteethand thenre measure Problem Cause Solution Erwil...

Страница 11: ...15 ApexConnectFileHolderA 815 1816 ApexConnectFileHolderB 815 1817 ApexConnectReplacementBattery ELECTROMAGNETIC COMPATIBILITY This device has been tested for compliance with electromagnetic tests acc...

Страница 12: ...errdental com Manufactured by Meta Systems Co Ltd 1214 18 Sicox tower 12F 484 Dunchon daero Jungwon gu Seongnam si Gyeonggi do 13229 Korea Distribution Facility Kerr Australia Pty Ltd 12 Mars Rd Unit...

Отзывы: