background image

STEP 4: INSTALL THE SIZING BARS

1. 

Using the table below, select the size of Sizing 

Bars best suited for the current participant.

Sizing Bar Measurement Chart 

Circumference

52-54cm

55-57cm 58-60cm

Sizing Bars

S (White)

M (Gray) L (Black)

TIP: Sizing Bars may be mixed to find the best 

possible fit. Notate the bars selected and use the 

same set every time that participant is recorded.

2. 

Turn the headset upside down and place it onto 

the inverted protective foam insert from the ship- 

ping case, tightening and loosening as described 

in the 

Handling the Headset

 section above. 

3. 

Insert the four Sizing Bars, using the magnets 

to snap the bars into the proper position as 

illustrated in the diagram below.

TIP: You may need to loosen the Tension Dial in 

order to install the Top Sizing Bar. 

STEP 5: PLACE THE HEADSET

1. 

Loosen the Tension Dial, then, holding the device 

by the Support Tabs, gently set the headset on 

the participant’s head.

2. 

Using two hands, carefully adjust the headset so 

the Alignment Guides on each side of the device 

align to matching positions on each side of the 

head.

3. 

Adjust the headset so the Alignment Guides on 

the front of the device are positioned symmetrical-

ly above the eyes. 

TIP: You may need to rotate the headset forward  

towards the brow for proper alignment.

4. 

Tighten the Tension Dial so the headset is as snug 

as possible while still remaining comfortable for 

the participant. 

STEP 6: TUNE THE LASERS

Once the headset is seated on the participant’s head, 

you must tune the lasers. This operation turns on the la-

sers and configures the strength of each laser to ensure 

every detector is receiving an adequately strong signal. 

NOTE: The Tune Lasers tool must be re-run each time 

the headset is placed on a participant’s head. 

1. 

In the Kernel Flow application, at the top of the 

main window, click the 

Tune Lasers

 button.

2. 

Click

 Tune Lasers

 in the window that appears.

A progress bar appears to indicate the duration of 

the operation.

3. 

Once the operation is complete, click 

Close

 to 

close the Tune Lasers window.

NOTE: When the Tune Lasers command is complete 

the lasers remain enabled. 

SAFETY WARNING:

 Never remove headset from 

the participant’s head while lasers are enabled.

8

7

Содержание Flow 50

Страница 1: ......

Страница 2: ...on is normally afforded by the aversion response The aversion response is defined as closure of the eyelid eye movement pupillary constriction or movement of the head to avoid an exposure to a noxious...

Страница 3: ...WHAT S IN THE BOX 2 1...

Страница 4: ...hipped in the future 1 Remove the foam tray containing the Module Storage Boxes and set it aside 2 Remove the Instapak foam insert covering the headset Do not discard this 3 Take the Flow headset out...

Страница 5: ...LIZE THE HEADSET 1 Be sure the headset is seated and tightened on the Headset Stand 2 In the main window of the Kernel Flow desktop application click Initialize 3 A confirmation dialog appears Click I...

Страница 6: ...n each side of the device align to matching positions on each side of the head 3 Adjust the headset so the Alignment Guides on the front of the device are positioned symmetrical ly above the eyes TIP...

Страница 7: ...STEP 8 RECORD A SESSION NOTE Remember that participants must explicitly provide consent by accepting a participant Invita tion Code before their session data may be viewed downloaded or analyzed 1 In...

Страница 8: ...Never remove the headset from a participant while the lasers are enabled 3 Loosen the Tension Dial and gently spread the plates apart then lifting from the Support Tabs carefully remove the headset fr...

Страница 9: ...2021 Kernel 090821 01...

Отзывы: