7
Display unit
7.1 Overview of displays
The three indicators provide the following information on the display unit:
◄
The weighing system (incl. load) is stable
▬
The indicated weight has a negative value
NET
◄
The indicated value is a net weight
The following messages may appear on the display:
HELP 1 Weighing system overloaded
HELP 2 Taring impossible due to negative gross weight
HELP 3 Negative signal of load cells to A-D converter / slant
HELP 4
Entered tare weight too high. To cancel the HELP display press
and
enter a new, reduced tare weight.
HELP 7 The signal of the load cells on the A-D converter is too high.
LO-BA
The rechargeable battery load status is too low; charge the
rechargeable battery.