8 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out during the initial start-up, after change in
location and variation of surrounding temperature. To receive accurate measuring
values it is also recommended to adjust the balance periodically in weighing
operation.
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization. Ensure that the system is unloaded. The system should be
lifted by approx. 2 lifting index marks and stand absolutely clear.
In verified weighing systems observe the check number.
After switching on the display unit the check number is displayed. This
number will be automatically adjusted after every interference. In verified
systems this number is an integral part for conformity evaluation. With that it
is always possible to check whether a new adjustment has been carried out.
In order to display the check number, switch off and switch on the system by
. The display unit shows subsequently e.g.:
Ö
8888.8 (check of LCD segments)
Ö
P 7.3 (software no.)
Ö
14325 (check number)
Ö
0 (weighing mode)
The verification is not valid, when the check number of the system does not
match with the check number on the type plate.
8.1 Zeroing
¾
Unload weighing system and switch on with
.
¾
Press
approx. 8 sec until the display counts reverse from AF08 to AF00,
followed by a determined percentage, e.g. AP 6.4. This percentage must not be
more than 20.
¾
After zeroing the system will automatically return into the standard weighing
mode.
VHM_M-BA-e-1012