10 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible
cause
The displayed weight does not glow.
•
The balance is not switched on.
•
The batteries are empty
The displayed weight is permanently
changing
•
Draught/air
movement
-
•
Sites with vibration.
•
The fork is in contact with foreign matter.
•
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing value is obviously
•
The display of the balance is not at zero
wrong
•
Adjustment is no longer correct.
•
Great fluctuations in temperature.
•
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
•
The balance is not on reference height.
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.