Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization. Ensure that there are no objects on the weighing pan.
Arrange adjustment weight, details see chap.1 „Techn. data“
•
Start balance by pressing
. During the selftest press
. The password query "P in" is displayed.
•
Using numeric keyboard, enter the standard password "9999".
"P in ----" is displayed.
•
Confirm by
, the menu is called up, the first menu point "F0
CAL" is displayed.
•
Press adjustment switch on the lower side of the balance
•
Press
, "F0 UnLoad" will be displayed. Ensure that there
are no weighing goods on the weighing pan. Wait for stability
display.
•
Press
anew. The value of the adjustment weight appears.
(example)
•
Either place the displayed adjustment weight, or use the numeric
keys to enter a new adjustment weight.
•
Acknowledge with
. „Load“ is displayed.
•
Place adjustment weight, wait for stability display and press
. “PASS“ will be displayed.
•
Adjustment now has finished. Remove adjustment weight from
weighing plate. The balance changes automatically into weighing
mode. The zero display appears.