RFE-BA-e-1811
15
7.7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
The adjustment is locked for verified balances. Carrying out adjustment
requires that the seal is destroyed and the adjusting switch is confirmed in
step 3 when turning on the scale. Position of adjustment switch, see chap.
7.8.1
Attention
:
After destruction of the seal the balance must be re-verified by an
authorised agency and a new verification wire/seal mark fitted before it can
be reused for applications subject to verification.