PES/PEJ-BA-defsi-0510
117
7.2.2. Parameter for serial interface
Not shown for menu setting „
v K.F 0
“ (interface de-activated).
Function
Display
or
Selection
or
Description of the
options
0
No data output
1
Continuous data output
2
Continuous data output stable weighing val-
ues
3
Output for stable and instable weighing val-
ues after pressing PRINT key
4
Output for stable weighing value after previ-
ous relief of balance
5
One output for stable weighing value. No
output for stable weighing values. Renewed
output after stabilization
6
One output for stable weighing value. Con-
tinuous output for instable weighing values.
*
7
Output of stable weighing values after press-
ing PRINT key
A
Single, immediate output after fixed interval
(see chpt. 14.5)
Output condition at
interface
61.
Q.m.
B
Single, immediate output after fixed interval
and stable weighing value (see chpt 14.5)
*
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
Baud rate
6r
B.L.
5
19200 bps
Содержание PEJ
Страница 3: ......