PES/PEJ-BA-defsi-0510
105
Carefully place adjusting weight in the centre
of the weighing plate
Adjustment process is started.
The process of adjustment is completed.
Remove adjusting weight, balance will return
into weighing mode automatically. In case of
an adjustment error or incorrect adjusting
weight the display will show
[
- Err
]
; repeat
adjustment process.
Qn
F.5.
CAL
!
Qn
F.5.
CAL
!
BWSX
!
EnD
CAL
!
a
z0
g
6.9.2. Adjustment test with external weight (only PES)
During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Procedure:
Observe stable environmental conditions. A warming up time of ca. 1 hour is required
for stabilization. Ensure that there are no objects on the weighing plate.
Operation
Display
Activate function
[
w CA.
4
]
(see chap. 7).
w CA 4
!
a
z0
g