5 Unpacking, Setup and Commissioning
5.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in
common conditions of use.
You will work accurately and fast, if you select the right location for your balance.
On the installation site observe the following:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing next to
a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time. Non-
permitted condensation (condensation of air humidity on the appliance) may
occur if a cold appliance is taken to a considerably warmer environment. In this
case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature.
Avoid static charge of goods to be weighed and weighing container.
Do not operate in areas with hazard of explosive material or in potentially
explosive atmospheres due to materials such as gasses, steams, mists or dusts.
Keep away chemicals (such as liquids or gasses), which could attack and
damage the balance inside or from outside.
Keep IP protection of the device.
If electro-magnetic fields or static charge occur, or if the power supply is unstable
major deviations on the display (incorrect weighing results) are possible. In that
case, the location must be changed.
5.2 Unpacking/installation
Carefully remove the balance from the packaging, remove plastic cover and setup
balance at the intended workstation.
Scope of delivery / serial accessories
Balance
Weighing plate
Mains adapter
Protective cover
Operating instructions
Windshield (only models PCD 400-3)