KERN OCO-2 Скачать руководство пользователя страница 6

4

6. Preparation for use

Please refer to diagram 1 on page 3 about the names 

of the following items.
6-1      Attaching the Eyepieces
6-1-1    Remove the eyepieces' dust caps.
6-1-2    Insert G10X eyepiece into the eyepiece sleeve.
                                                                                                                                               

6-2      Attaching the Objectives

6-2-1 Remove the stage center plate and attach the
objectives to the revolving nosepiece through the hole
on the stage left by the plate.
6-2-2      Attaching the objectives in such a manner that
  the magnification increase from low to higher powers
in the clockwise direction.
6-2-3    In the inverted microscope, the front lens of the
objectives faces upward, and is more exposed to
contamination than the objectives of upright
  microscopes. Therefore, if there are empty positions

6-3    Attaching the fixator of the universal handwheel

6-3-1      This is an optional part which can be attached
according to your demand.
6-3-2 Joint the fixator to the stage with hexagonal
screws on , and to the connecting-rod of the

6-4      Attaching the Stage

6-4-1 Gently place the stage on the microscope frame by 
aligning  the  stage  mounting  holes  with  the threaded 
holes on the frame.
6-4-2 Insert the provided Allen screws into the mounting 
holes. Tighten the screws using the
provided Allen wrench.
6-4-3 The stage is designed very thin so that the objective 
will not hit it when the revolving nosepiece is rotated. Do 
not subject the stage centeplage to impact or excessive force, as this may deform it.

6-1

6-2

6-3

6-4

Содержание OCO-2

Страница 1: ...Operating Manual Inverted System Microscopes KERN OCO 2...

Страница 2: ...f the text symbols 2 4 Precautions When Unpacking the Microscope 2 5 Module Nomenclature See diagram 1 3 6 Preparation for use 4 7 Preparations for the use of special devices 6 8 Operations 7 9 Troubl...

Страница 3: ...ve precipitation and crystal structure analysis The microscopes are used in many practical areas as well such as bio medical science preventive medication environmental engineering food processing pha...

Страница 4: ...the Revolving Nosepiece Never attempt to rotate the coarse or fine adjustment knob without removing the clamping rod Otherwise the focusing mechanism may be damaged 1 Loosen he screw of the clamping r...

Страница 5: ...video port C mountdigitalport C mountvideo Camera three position slider reflector DSZ main body Analyzer UCIS objective PH UCIS objective specimen stage re ctangular stage With flexible right handle a...

Страница 6: ...ed to contamination than the objectives of upright microscopes Therefore if there are empty positions 6 3 Attaching the fixator of the universal handwheel 6 3 1 This is an optional part which can be a...

Страница 7: ...proper cord is provided If the power outlet is not grounded properly we can no longer warrant the electrical safety performance of the equipment 6 6 4 If the power cord or a connection cable comes in...

Страница 8: ...ide port is not used attach the cap for protecting it from dust 6 10 Attaching the TV Observation System 6 10 1 This is an optional part for TV observation 6 10 2 Attaching the photographic eyepiece a...

Страница 9: ...ght path Select the light path with the light path selector lever Place the specimen on the stage Engage a 10X objective in the light path Bring the specimen into focus Adjust the brightness Adjust th...

Страница 10: ...10Xobjective in the light path 8 2 6 Bring the specimen into focus 1 Rotation Direction of the Coarse Fine Adjustment Knobs Rotating the coarse or fine adjustment knob toward the front in the directi...

Страница 11: ...knob The stage up down movement using the fine adjustment knob is not limited When the pre focusing lever is locked the coarse adjustment stroke is limited by the mechanism and it cannot reach the pr...

Страница 12: ...ver 8 3 4 Attach the phase contrast slider 8 3 5 Engage the objective matched with the phase contrast slider in the light path 8 3 6 Disengage the 10X eyepiece and insert the WY1 central telescope 8 3...

Страница 13: ...ightness 8 3 13 Adjust the interpupillary distance and diopter 8 3 14 Engage the objective to be used in the light path and bring the specimen in focus 8 3 15 Adjust the aperture iris and field iris d...

Страница 14: ...ed Reduce the filters to the minimum required b The bulb lights but the field of view is dark The stage central plate is engaged in the optical path Move the stage and place the specimen again Light p...

Страница 15: ...asily or at all The pre focusing lever is locked Unlock it j The revolving nosepiece lowers by its own weight or defocusing occurs due to slipping of fine adjustment The tension adjustment ring is too...

Страница 16: ...ure for cleaning for these will deform the electrically insulated section of the extremity Since solvents such as ether and alcohol are highly flammable they must be handled carefully Be sure to keep...

Отзывы: