14

power  cord  from  the  wall  outlet,  then  allow  the  old  bulb  and  lamp  socket  to  cool  before 
replacing the bulb with a new of the designated type.
2 Attaching the Lamp Socket
Insert the plug into the socket, then push the guide pins gently into the guide holes.

10. Maintenance of the microscope

1.    Clean all  glass components by  wiping gently  with  gauze.  To  remove  fingerprints  or  oil 

smudges, wipe with gauze slightly moistened with a mixture of ether(75%) and alcohol(25%)
        To clean the extremity of an immersion objective, use neutral detergent. Do not use the 
ether/alcohol mixture for cleaning , for these will deform the electrically insulated section of the 
extremity.
        Since  solvents  such  as  ether  and  alcohol  are  highly  flammable,  they  must  be  handled 
carefully.Be  sure  to  keep  these  chemical  away  from  open  flames  or  potential  sources  of 
electrical  sparks  -  for  example,  electrical  equipment  that  is  being  switched  on  or  off.  Also 
remember to always use these chemical only in a well-ventilated room.
2.  Be  sure  to  clean  the  oil  immersion  objective  after  use. Leaving  immersion  oil  on  it  will 
degrades its performance.
3. Do not attempt to disassemble any part of the microscope.
4. When not using the microscope, make sure to set the main switch to "O"(OFF), confirm that 
the lamp housing is cool enough and cover the microscope with the provided dust cover.

Содержание OCO-2

Страница 1: ...Operating Manual Inverted System Microscopes KERN OCO 2...

Страница 2: ...f the text symbols 2 4 Precautions When Unpacking the Microscope 2 5 Module Nomenclature See diagram 1 3 6 Preparation for use 4 7 Preparations for the use of special devices 6 8 Operations 7 9 Troubl...

Страница 3: ...ve precipitation and crystal structure analysis The microscopes are used in many practical areas as well such as bio medical science preventive medication environmental engineering food processing pha...

Страница 4: ...the Revolving Nosepiece Never attempt to rotate the coarse or fine adjustment knob without removing the clamping rod Otherwise the focusing mechanism may be damaged 1 Loosen he screw of the clamping r...

Страница 5: ...video port C mountdigitalport C mountvideo Camera three position slider reflector DSZ main body Analyzer UCIS objective PH UCIS objective specimen stage re ctangular stage With flexible right handle a...

Страница 6: ...ed to contamination than the objectives of upright microscopes Therefore if there are empty positions 6 3 Attaching the fixator of the universal handwheel 6 3 1 This is an optional part which can be a...

Страница 7: ...proper cord is provided If the power outlet is not grounded properly we can no longer warrant the electrical safety performance of the equipment 6 6 4 If the power cord or a connection cable comes in...

Страница 8: ...ide port is not used attach the cap for protecting it from dust 6 10 Attaching the TV Observation System 6 10 1 This is an optional part for TV observation 6 10 2 Attaching the photographic eyepiece a...

Страница 9: ...ght path Select the light path with the light path selector lever Place the specimen on the stage Engage a 10X objective in the light path Bring the specimen into focus Adjust the brightness Adjust th...

Страница 10: ...10Xobjective in the light path 8 2 6 Bring the specimen into focus 1 Rotation Direction of the Coarse Fine Adjustment Knobs Rotating the coarse or fine adjustment knob toward the front in the directi...

Страница 11: ...knob The stage up down movement using the fine adjustment knob is not limited When the pre focusing lever is locked the coarse adjustment stroke is limited by the mechanism and it cannot reach the pr...

Страница 12: ...ver 8 3 4 Attach the phase contrast slider 8 3 5 Engage the objective matched with the phase contrast slider in the light path 8 3 6 Disengage the 10X eyepiece and insert the WY1 central telescope 8 3...

Страница 13: ...ightness 8 3 13 Adjust the interpupillary distance and diopter 8 3 14 Engage the objective to be used in the light path and bring the specimen in focus 8 3 15 Adjust the aperture iris and field iris d...

Страница 14: ...ed Reduce the filters to the minimum required b The bulb lights but the field of view is dark The stage central plate is engaged in the optical path Move the stage and place the specimen again Light p...

Страница 15: ...asily or at all The pre focusing lever is locked Unlock it j The revolving nosepiece lowers by its own weight or defocusing occurs due to slipping of fine adjustment The tension adjustment ring is too...

Страница 16: ...ure for cleaning for these will deform the electrically insulated section of the extremity Since solvents such as ether and alcohol are highly flammable they must be handled carefully Be sure to keep...

Отзывы: