5.6  Adjusting the illumination 

 
To  make  sure  that  perfect  image  results  are  achieved  during  microscopic 
observation, it is important that the direction of light of the microscope is optimised.  
 
The  necessary  control  elements  for  this  are  the  height-adjustable  condenser  with 
aperture diaphragm. 
 
 
When adjusting the lighting for the first time, you must first select the lowest possible 
objective magnification, so that you can carry out the following steps. 
 
 
 
 
1.  Adjust  the  height  of  the  condenser  by  turning  the  condenser  focus  dial  to  get  a 

good contrast of the microscopic image. Normally therefore you have to bring the 
condenser to just below the maximum height. 

 
 
 
2.  Use  the  aperture  diaphragm  of  the  condenser  to  find  the 

very best compromise between contrast and resolution for 
the  microscopic  image.  For  the  objective  with  the  lowest 
magnification  the  lever  of  the  aperture  diaphragm  should 
be placed almost completely on the right-side limit, so that 
the opening of the diaphragm is very small. The higher the 
magnification of an objective, the larger the opening should 
be selected by pushing the lever towards the left-side limit.  

 
 

The  view  in  the  tube  without  the  eyepiece  should  look 
something like the illustration on the right. 
The  diameter  of  the  aperture  diaphragm  which  is  then  visible  should  make  up 
approximately 2/3 of the pupil diameter. 

 

If the eyepiece should be removed, for checking, then please make sure that no 
dirt or dust falls into the tube. 

 
 
 
3.  The brightness is always controlled by the bulb brightness (using the dimmer) and 

not by the aperture diaphragm. 

 
 
 
 

 

Содержание OBE 101

Страница 1: ...kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Transmitted light laboratory microscope KERN OBE 1 OBE 101 OBE 102 OBE 103 OBE 107 OBE 108 OBE 1...

Страница 2: ......

Страница 3: ...es 11 4 4 Condenser 11 5 Operation 12 5 1 Getting started 12 5 2 Pre focussing 13 5 3 Adjusting the interpupillary distance 14 5 4 Dioptre adjustment 14 5 5 Adjusting the magnification 15 5 6 Adjustin...

Страница 4: ...ctric shock 1 2 Notes on the electrical system Before connecting to a mains power supply you must make sure that you are using the correct input voltage The information to select the correct mains cab...

Страница 5: ...microscope is not being used you should cover it with the enclosed dust protective cover When doing this the power supply is stopped by switching off at the main switch and unplugging the mains cable...

Страница 6: ...these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However organi...

Страница 7: ...OBE 1 BA e 1611 6 2 Nomenclature...

Страница 8: ...7 OBE 1 BA e 1611 Rear view...

Страница 9: ...ocular WF 10x 18 mm Achromatic 4x 10x 20x 40x 3W LED Transmitted OBE 109 Finite Binocular WF 10x 18 mm Achromatic 4x 10x 20x 40x 3W LED Transmitted Accu OBE 110 Finite Trinocular WF 10x 18 mm Achromat...

Страница 10: ...9 OBE 1 BA e 1611...

Страница 11: ...OBE 1 BA e 1611 10...

Страница 12: ...agnification for both eyes These are already placed onto the tube connectors and each is fixed with a small silver screw under the eyepiece on the tube connectors so that they are turnable but you can...

Страница 13: ...time your eyes are not immediately subject to a high level of light The next step is to place an object holder with the sample on the angle table To do this the cover glass must be facing upwards You...

Страница 14: ...soon as an image is recognisable no matter how sharp then you should only adjust the focus using the fine adjustment knob Adjusting the torque of the coarse and fine adjustment knob Next to the left a...

Страница 15: ...cular image then the interpupillary distance is set correctly 5 4 Dioptre adjustment For binocular and trinocular devices The eye strengths of each eye of the microscope user can often be slightly dif...

Страница 16: ...objectives you should use the silver ring above the objectives see illustration When rotating the nosepiece you must always make sure that the objective which is about to be positioned in the beam pat...

Страница 17: ...ure diaphragm of the condenser to find the very best compromise between contrast and resolution for the microscopic image For the objective with the lowest magnification the lever of the aperture diap...

Страница 18: ...ries Battery type LR6 AA Capacity 1300 mAh Nominal voltage 1 2 V Charging time ca 5 h Operation time with ideal light intensity ca 3 h Total operation time ca 8 h The batteries are being charged as so...

Страница 19: ...sses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved We would theref...

Страница 20: ...Lower the specimen stage and position the 100x objective in the beam path 3 Bring the specimen stage or object slide to the objective very slowly until there is slight contact 4 Observe the object The...

Страница 21: ...nection point of its cable Now the new module has to be mounted in the same why as the original one After the bulb holder has been replaced in the underside of the device and the screws replaced the b...

Страница 22: ...wo points to take into account when subsequently using the polarisation unit The aperture setting of the condenser must be at the maximum lever completely on the left side For its starting position th...

Страница 23: ...ew A camera which has a C mount thread is then screwed on top of the adapter We recommend that you first adjust the field of view using the eyepieces on the device for the existing requirements and th...

Страница 24: ...y centred The condenser is too low The field of view is dark or is not correctly illuminated The objective is not positioned correctly on the beam path The selector switch for the beam path is between...

Страница 25: ...Dirt dust on the front lens of the condenser Dirt dust on the object One side of the image is blurred The stage was not correctly fitted The objective is not positioned correctly on the beam path The...

Страница 26: ...materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...

Отзывы: