11 

 

OBE-1-BA-e-1611 

Assembly 

 

4.1  Microscope head 

 
Inside  the  packaging  the  microscope  head  is  already  mounted  but  inclined  towards 
the rear. In order to turn it to the front you must loosen the fixing screw on the tube 
connection point and after the turning fix it again with the screw. In case of removing 
the head completely from the housing, you should always make sure that you do not 
touch the lenses with your bare fingers and that no dust enters the apertures. 
 
 

4.2  Objectives 

 
All  four  objectives  are  already  mounted  to  the  nosepiece.  After  removing  the 
protective foil they are ready for use. They are ranged in such a way that if you turn 
the  nosepiece  clockwise,  the  objective  with  the  next  higher  magnification  appears. 
When the objectives need to be dismounted, you should always make sure that you 
do not touch the lenses with your bare fingers and that no dust enters the apertures. 
For  objectives  which  are  marked  “OIL”,  you  must  use  an  immersion  oil  with  the 
lowest level of inherent fluorescence. 
 
 

4.3  Eyepieces 

 
You  must  always  use  eyepieces  with  the  same  magnification  for  both  eyes.  These 
are  already  placed  onto  the  tube  connectors  and  each  is  fixed  with  a  small  silver 
screw under the eyepiece on the tube connectors, so that they are turnable but you 
cannot  pull  them  out  of  the  tube.  When  changing  the  eyepieces,  you  must  loosen 
these  screws  and  once  the  new  eyepiece  is  fitted,  then  they  must  be  re-tightened. 
You should always make sure that you do not touch the lenses with your bare fingers 
and that no dust enters the apertures. 
 
 

4.4  Condenser 

 
The condenser is firmly fixed onto a holding ring (condenser holder) underneath the 
microscope stage. The lever for the aperture diaphragm is directed towards the front. 
There  is  the  ability  to  adjust  the  height  of  the  condenser,  but  not  to  centre  it.  We 
recommend that you use the course adjustment knob to bring the specimen stage to 
its uppermost position when you need to remove the condenser. Then use the focus 
dial of the condenser to move the condenser holder to a low position. In this way the 
condenser can be taken off after loosening the three screws on the holding ring.  

If the microscope has no focus dial for the condenser, the height adjustment is 
done by turning the condenser on its vertical axis. 

You should avoid touching the optical lenses with bare fingers. 
 

 

Содержание OBE 101

Страница 1: ...kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Transmitted light laboratory microscope KERN OBE 1 OBE 101 OBE 102 OBE 103 OBE 107 OBE 108 OBE 1...

Страница 2: ......

Страница 3: ...es 11 4 4 Condenser 11 5 Operation 12 5 1 Getting started 12 5 2 Pre focussing 13 5 3 Adjusting the interpupillary distance 14 5 4 Dioptre adjustment 14 5 5 Adjusting the magnification 15 5 6 Adjustin...

Страница 4: ...ctric shock 1 2 Notes on the electrical system Before connecting to a mains power supply you must make sure that you are using the correct input voltage The information to select the correct mains cab...

Страница 5: ...microscope is not being used you should cover it with the enclosed dust protective cover When doing this the power supply is stopped by switching off at the main switch and unplugging the mains cable...

Страница 6: ...these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However organi...

Страница 7: ...OBE 1 BA e 1611 6 2 Nomenclature...

Страница 8: ...7 OBE 1 BA e 1611 Rear view...

Страница 9: ...ocular WF 10x 18 mm Achromatic 4x 10x 20x 40x 3W LED Transmitted OBE 109 Finite Binocular WF 10x 18 mm Achromatic 4x 10x 20x 40x 3W LED Transmitted Accu OBE 110 Finite Trinocular WF 10x 18 mm Achromat...

Страница 10: ...9 OBE 1 BA e 1611...

Страница 11: ...OBE 1 BA e 1611 10...

Страница 12: ...agnification for both eyes These are already placed onto the tube connectors and each is fixed with a small silver screw under the eyepiece on the tube connectors so that they are turnable but you can...

Страница 13: ...time your eyes are not immediately subject to a high level of light The next step is to place an object holder with the sample on the angle table To do this the cover glass must be facing upwards You...

Страница 14: ...soon as an image is recognisable no matter how sharp then you should only adjust the focus using the fine adjustment knob Adjusting the torque of the coarse and fine adjustment knob Next to the left a...

Страница 15: ...cular image then the interpupillary distance is set correctly 5 4 Dioptre adjustment For binocular and trinocular devices The eye strengths of each eye of the microscope user can often be slightly dif...

Страница 16: ...objectives you should use the silver ring above the objectives see illustration When rotating the nosepiece you must always make sure that the objective which is about to be positioned in the beam pat...

Страница 17: ...ure diaphragm of the condenser to find the very best compromise between contrast and resolution for the microscopic image For the objective with the lowest magnification the lever of the aperture diap...

Страница 18: ...ries Battery type LR6 AA Capacity 1300 mAh Nominal voltage 1 2 V Charging time ca 5 h Operation time with ideal light intensity ca 3 h Total operation time ca 8 h The batteries are being charged as so...

Страница 19: ...sses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved We would theref...

Страница 20: ...Lower the specimen stage and position the 100x objective in the beam path 3 Bring the specimen stage or object slide to the objective very slowly until there is slight contact 4 Observe the object The...

Страница 21: ...nection point of its cable Now the new module has to be mounted in the same why as the original one After the bulb holder has been replaced in the underside of the device and the screws replaced the b...

Страница 22: ...wo points to take into account when subsequently using the polarisation unit The aperture setting of the condenser must be at the maximum lever completely on the left side For its starting position th...

Страница 23: ...ew A camera which has a C mount thread is then screwed on top of the adapter We recommend that you first adjust the field of view using the eyepieces on the device for the existing requirements and th...

Страница 24: ...y centred The condenser is too low The field of view is dark or is not correctly illuminated The objective is not positioned correctly on the beam path The selector switch for the beam path is between...

Страница 25: ...Dirt dust on the front lens of the condenser Dirt dust on the object One side of the image is blurred The stage was not correctly fitted The objective is not positioned correctly on the beam path The...

Страница 26: ...materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...

Отзывы: