Troubleshooting 

 

Unit does not turn on 

Check mains cable 
Check fuses 
Check cable connections 
Problem on power board 
Problem on display board 
Problem on main board 

Balance is weighing but is 
unstable 

Mechanical damage 
Pan rubbing against case or not installed correctly 
Air drafts or vibration or unstable table 
Not good parameters for the working conditions 
Dirt inside the measurement system or inside the coil 

No zero point after 
turning on the balance 

Pan needs to zero 
Mechanical damage 

4 Balance 

shows 

wrong 

weight 

Wrong calibration 
Mechanical damage to weighing mechanism 
Pan not installed correctly 
Off center load error 
Dirt inside the measurement system 

Off center load error 

Check if the horizontal stat is being 
kept mechanical damage 

Intense creeping (drift) 

Mechanical damage 
Dirt inside the measurement system 

Linearity error 

Linearity correction 
Mechanical damage 

9 No 

communication 

between balance and PC 

Wrong parameter 
Check connection RS-232 to PC 
Wrong interface cable 

10 

Time and date does not 
work 

Check voltage TTL 2 Hz (0/+5V DC on measuring point 

SQW

 

Problem on main board 
Batteries are damaged 

11 

Display shows program 
number, no function 

Quartz does not work 
Processor damage 
Check tension 

RST

 (reset) 

“1” 

+5V and exchange main board 

when RST does not work 

12 

Outside maximum range 

Mechanical damage 
Check voltage on measuring resistor 

13 

Balance works OK but 
drying  is not correct 

Check parameters and mode selected 
Check samples are not hitting the temp. sensor 
Check temperature of chamber 
Check lamps 

14 Temperature 

of 

chamber 

is incorrect 

Recalibrate temperature sensor 
Temperature sensor damaged 
Lamp circuits damaged 

15 

Heating lamps will not 
turn on at any time 

Check lamps 
Check wiring of lamps 
Temperature sensor 

16 

Lamps will not turn off 

Damaged temp. sensor 
Recalibrate temp. sensor 
Lamp control circuits damaged 

 
 
 
 
 
 

4  

MLS…N-SH-e-0812 

Содержание MLS 50-3...N series

Страница 1: ...lei 1 D 72336 Balingen Frommern E Mail info kern sohn com Tel 0049 0 7433 9933 0 Fax 0049 0 7433 9933 149 Internet www kern sohn com GB Service Manual Electronic Moisture Analyzer Page 2 KERN MLS 50 3 N Version 1 2 10 2008 ...

Страница 2: ...ription of factory parameters 7 5 Factory Calibration 9 6 Temperature Calibration 10 7 Linearity Correction 11 7 1 Linearity correction single stage 11 7 2 Parabolic linearity correction 12 7 3 Linearity correction multi stage 13 7 4 Changing linearity factors 14 8 Off Center Load Adjustment 15 9 Printout of Factory Parameters 16 2 MLS N SH e 0812 ...

Страница 3: ...rying process Start Stop drying Cancel an entry Leave menu Arrow keys for navigation in menu Change the value of a parameter Data export to external device Confirm save settings Tara key scales zero digits Invoke user menu drying parameter setting Invoke user menu MLS N SH e 0812 3 ...

Страница 4: ... Linearity error Linearity correction Mechanical damage 9 No communication between balance and PC Wrong parameter Check connection RS 232 to PC Wrong interface cable 10 Time and date does not work Check voltage TTL 2 Hz 0 5V DC on measuring point SQW Problem on main board Batteries are damaged 11 Display shows program number no function Quartz does not work Processor damage Check tension RST reset...

Страница 5: ...ge Result 10 Max 2 7 Result 10 Result 4 Max 2 8 Load on weighing plate too heavy Difference 1 Max 2 9 Difference between cal weight stored current cal weight 1 Sample mass 20 mg 2 10 Sample 20 mg Sample mass above the range 2 11 Sample out of set range Above the range 3 1 Value of parameter outside Faulty value 3 2 Value of parameter outside Incorrect password 3 4 Wrong password Error of notice 4 ...

Страница 6: ... are parameters needed to start settings and regulations 4 1 How to enter the service menu Open the housing and set jumper JP1 on main board 172xxx PCB to access the service menu Plug the balance into the power outlet and enter service menu 6 MLS N SH e 0812 ...

Страница 7: ... 0d Normal 12 3 0 d P0 07 Autozero delay Time of autozero possibility to change value from 0 2s to 3 0s Normal 6 2 0 s P0 08 Stable range Stable range of measuring results possibility to change value from 0 1d to 10 0d Normal 11 2 0 d P0 09 Stable speed Time to define stable measuring results possibility to change value from 0 2s to 3 0s Normal 6 2 0 s P0 10 Filter range Digital filter possibility...

Страница 8: ...mperature compensation after filtering 2 nd temperature point P0 32 2 nd point T Second temperature correction set by 28 C P0 33 Temp Z factor Temperature compensation factor of zero P0 34 Temp S factor Temperature compensation factor of sensitivity P0 35 Lin points Lin value 1 Lin factor 1 Lin value 2 Lin factor 2 Lin value 3 Lin factor 3 P0 64 Lin value 15 P0 65 Lin factor 15 P0 66 Lin value 16 ...

Страница 9: ... moisture analyzer 0 Halogen 1 IR P0 75 Cor factor T Service parameter for characteristic of drying chamber thermometer 5 Factory Calibration Enter submenu P0 Factory and press key to start factory calibration P0 14 Return to weighing mode Press the ESC key repeatedly until the query SAVE appears Confirm query by pressing the PRINT key or reject it by pressing the ESC key MLS N SH e 0812 9 ...

Страница 10: ...red by using the arrow keys Ð Ï Setup 29 01 08 Setup Temperature calibration 1 point 00 00 Setup 29 01 08 Setup Temperature calibration Set temp value C 25 Setup 29 01 08 Setup Temperature calibration Set temp value C 24 To import the temperature values press the PRINT key temperature calibration for the second point will be started Setup 29 01 08 Setup Temperature calibration 2 point 14 59 Temper...

Страница 11: ...e Enter submenu P0 Factory Set index next to parameter P0 16 and press the Î button Return to weighing mode Press the ESC key repeatedly until the query SAVE appears Confirm query by pressing the PRINT key or reject it by pressing the ESC key MLS N SH e 0812 11 ...

Страница 12: ...nu P0 Factory Set index next to parameter P0 16 and press the Î button Return to weighing mode Press the ESC key repeatedly until the query SAVE appears Confirm query by pressing the PRINT key or reject it by pressing the ESC key 12 MLS N SH e 0812 ...

Страница 13: ...enu P0 Factory Set index next to parameter P0 16 and press the Î button Return to weighing mode Press the ESC key repeatedly until the query SAVE appears Confirm query by pressing the PRINT key or reject it by pressing the ESC key MLS N SH e 0812 13 ...

Страница 14: ...ence to the factor Æ changing in the factory menu service menu parameters P0 37 until P0 67 E g weight mass 40g Æ reading 40 007 stored linearity factor at 40g Æ 0 001 input new linearity factor at 40g Æ 0 008 weight mass 40g Æ reading 39 995 stored linearity factor at 40g Æ 0 001 input new linearity factor at 40g Æ 0 004 5 After inscribing correction you should do calibration again 6 Check balanc...

Страница 15: ... load at the centre than at the left rear right and front side of the pan of the way out from the centre The readings should agree within 5mg Using the attached figure It is necessary to change the screws to make the readings of the weight equal as it is moved around the pan Recheck corner load after 1 hour again MLS N SH e 0812 15 ...

Страница 16: ...FF key main switch to turn on the balance Enter main menu Set index next to parameter P7 Globals P1 alibration P2 GLP P3 D P5 RS 232 P6 P7 C ate Time P4 Redout Printouts Globals Setup 29 12 04 Enter menu P7 Globals Enter submenu P7 09 Par printout To get a printout of the factory parameters P0 01 to P0 75 set Start 0 01 Stop 0 75 Confirm your setup with PRINT Press PRINT to start the printout 16 M...

Отзывы: