7.4
Changing linearity factors
1. After come back to weighing mode do calibration.
2. Check balance readings in all measuring points.
3. In case of differences make a note.
4. If there is a positive difference you should add this difference to the factor.
If there is a negative difference you should subtract this difference to the factor.
Æ
changing in the factory menu (service menu), parameters
P0 37
until
P0 67
E.g.:
weight mass 40g
Æ
reading 40.007
stored linearity factor at 40g
Æ
0.001
input
new linearity factor
at 40g
Æ
0.008
weight mass 40g
Æ
reading 39.995
stored linearity factor at 40g
Æ
0.001
input
new linearity factor
at 40g
Æ
- 0.004
5. After inscribing correction you should do calibration again.
6. Check balance readings in all measuring points.
7. Repeat this process until you will get the expected results.
14
MLS…N-SH-e-0812