KMF-TM-BA_IA-e-1310
33
Procedure:
In weighing mode, press
repeatedly until
[tCH]
appears.
Press
,
[Pin]
is displayed
Press
,
and
one after the other,
[P1 SPd]
will
appear
Press
, „
P2 CAL
“ will be displayed
Operate adjustment switch; for position see chap.13
Press
,
[dESC]
is displayed
Press
repeatedly until
[CAL]
will be displayed.
Confirm with
,
[UnloAd]
appears
Ensure that there are no objects on the weighing pan.
Wait for stability display "STABLE", then confirm with
(example)
The size of the currently set adjustment weight will be
displayed. The active digit is flashing.
To change, select the digit to be altered by
, and the
numerical value by
.
Confirm with
,
[LoAd]
appears.
Содержание KMF-TM
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 133: ......
Страница 134: ......
Страница 135: ......
Страница 136: ......
Страница 137: ......
Страница 138: ......
Страница 139: ......
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......