KFE-TNM-BA_IA-e-1610
29
9.4 Error messages
Error
message
Description
Possible causes
- - - - -
Maximum load exceeded
•
Unload weighing system or reduce
preload.
- - ol - -
Err 1
Incorrect data input
•
Follow format “yy:mm:dd“
Err 2
Incorrect time entry
•
Follow format “hh:mm:ss“
Err 4
Zeroing range exceeded due
to switching-on balance or
pressing
(normally 4% max)
•
Object on the weighing plate
•
Overload when zeroing
Err 5
Keyboard error
Err 6
Value outside the A/D
changer range
•
Weighing plate not installed
•
Damaged weighing cell
•
Damaged electronics
Err 9
Stability display does not
appear
•
Check the environmental
conditions.
Err 17
Taring range exceeded
•
Reduce load
Fai l h /
Fai l l
Adjustment error
•
Repeat adjustment.
Ba lo /
Lo ba
Battery very low
•
Recharge battery
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.