KFB/KFN-TAM-BA_IA-e-1832
27
7.7 Weighing with tolerance range
You can set an upper or lower limit when weighing with tolerance range and thus
ensure that the weighed load remains exactly within the set limits.
During tolerance tests such as dosing, portioning and sorting the unit will indicate
exceeded or undershot limits by emitting an optical or acoustic signal.
Audio signal:
The acoustic signal depends on the settings in menu block „BEEP“.
Options:
•
no
Acoustic signal turned off
•
ok
An acoustic signal sounds when load is within tolerance
limits
•
ng
An acoustic signal sounds when load is beyond tolerance
limits
Optical signal:
Three colour signal lights indicate whether the load is within the two tolerance limits.
The signal lights provide the following information:
+
Goods to be weighed above tolerance limit
Red signal light
glowing
Goods to be weighed within tolerance range
Green signal light
glowing
-
Goods to be weighed below tolerance limit
Red signal light
glowing
The settings for tolerance weighing may be called up either via menu block „
P0 CHK
“
(see chap. 8) or faster via the key combination