90
ILJ-C/ILS-C-BA-e-1320
17.1.1 Setting an adjustment report
The following settings can be enabled (printout takes place) or disabled (printout
blocked) by selecting YES or NO.
Sample adjustment record:
Select P 5.1 and adjust the desired settings.
Sample adjustment record:
P5.1.1
DRAFT
Enter a project name, maximum 16 characters
P5.1.2
ART ADJUST
YES
Print kind of adjustment
P5.1.3
USER
YES
Print name of user
P5.1.4
DRAFT
YES
Print project name (See. P5.1.1)
P5.1.5
DATUM
YES
Print date of adjustment
P5.1.6
TIME
YES
Print time of adjustment
P5.1.7
ID BALANCE
YES
Print serial number of weighing scale
P5.1.8
DIFF.ADJUST.
YES
Print difference between currently adjustment
weight and previous adjustment weight.
P5.1.9
LINES
YES
Print line between date and signature
P5.1.10 SIGNATURE
YES
Print box for signature of implementing person
Printout example adjustment report:
-------------------Just.Prot.------------------
Art Just
External
User
Kern
Draft
Date
21.11.2013
Time
9:54:39
ID Balance
132051
Diff.Adjust.
-0.2 g
Signature
……………………………………………..
Содержание ILJ 25K-4CM