HFM-BA-e-1813
26
8. Error messages
Error
message
Description
Possible causes
--ol-
Maximum load exceeded
Reduce load
Check whether the balance has
been damaged
Err 5
Keyboard error
Improper operation of the balance
Err 6
Value outside the A/D changer
range
Weighing plate not installed
Damaged weighing cell
Damaged electronics
Ba lo
Capacity of rechargeable
battery exhausted
Recharge battery
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
Содержание HFM 10T1
Страница 21: ...HFM BA d 1813 20 Nicht ziehen oder schieben Nicht seitlich am Haken ziehen Keine Mehrfachaufh ngungen verwenden...
Страница 33: ...HFM BA d 1813 32...
Страница 36: ...HFM BA d 1813 35 Kranwaage Modell Serien Nr Teil Ma nahme Datum Name Unterschrift...
Страница 37: ......
Страница 56: ...HFM BA e 1813 20 Do not push or pull Do not pull the hook horizontally Do not use multiple suspensions...
Страница 67: ...HFM BA e 1813 31...
Страница 70: ...HFM BA e 1813 34 Crane scales Model Serial no Part Action Date Name Signature...
Страница 71: ......
Страница 101: ...HFM BA f 1813 31...
Страница 104: ...HFM BA f 1813 34 Balance grue Mod le N de s rie Pi ce constitutive Mesure Date Nom Signature...