EOB / EOE / EOS -BA-e-1610
27
In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.
Help:
Fault
Possible cause
The displayed weight does not glow.
•
The balance is not switched on.
•
The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
•
Power supply interrupted.
The displayed weight is permanently
changing
•
Draught/air movement
•
Table/floor vibrations
•
The weighing pan is in contact with
foreign matter.
•
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing value is obviously
wrong
•
The display of the balance is not at zero
•
Adjustment is no longer correct.
•
The balance is on an uneven surface.
•
Great fluctuations in temperature.
•
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.