30
EC-BA-defs-0113
Error messages
b-Err
Electronic error by a static electricity or noise.
Contact the shipper.
L-Err
In counting the unit weight of samples is too
light for the balance division. Countable unit
weight is the readability of the balance or over.
o-Err
The load exceeds the capacity of the balance.
The tare is too heavy.
u-Err
Something contacts the weighing pan to lift it.
1-Err
In span
Reference weight is less than ½ FS.
2-Err
adjusting
Error exceeds 1%.