14  Important Information 

 
These electronic balances are precision instruments. It is designed for use in a self-
controlled electromagnetic environment. That means that in such an environment 
radio equipment, such as mobile phones, must not be used in direct vicinity. In case 
strong electromagnetic fields should arise, deviations of the indication are possible. 
Avoid disturbing environmental conditions such as currents and vibrations as well as 
a rapid change of the ambient temperature. If necessary readjust the balance after a 
change of temperature. 
 
Avoid high humidity, vapours and dust as the balance is not hermetically closed. Do 
not let the balance come into direct contact with liquids as these could enter the 
measuring cell. Therefore only clean the balance with a dry or damp cloth. Do not 
use solvents as varnished or plastic parts could be damaged. 
 
A warming up time of a few minutes after switching the balance on will stabilise the 
measuring values. Place the object carefully on the weighing pan. Avoid any constant 
load on the weighing pan even when out of use. Avoid shocks, thrusts and extreme 
impairments by all means. Remove spillages immediately. 
 
When the programme cycle is disturbed switch the balance off and then on again. 
Start the weighing operation again. 
 
Never use the balance in rooms where there is a danger of explosion. The serial 
version is not explosion-proof. 
 
The balance should be adjusted regularly with the given adjusting weight. 
 
Keep the packing materials for any possible reshipment. When reshipping, use only 
the original packing materials. 
 

Opening of the balance causes invalidity of warranty. 

Содержание EC 12K 1

Страница 1: ...ON TO KERN EC COUNTING BALANCES 19 3 UNPACKING AND ACCESSORIES 19 4 SETTING UP 19 5 LEVELLING 20 6 ELECTRICAL SUPPLY 20 7 DECLARATION OF CONFORMITY 21 8 EXTERNAL VIEW NAMES OF PARTS 22 9 INSTALLATION...

Страница 2: ...ty 0 01 g 0 1 g 1 g Linearity 0 02 g 0 2 g 2 g Adjustment weight 300 g M1 3000 g M1 5000 g M2 Weighing Platform 140 mm 190 x 190 mm Dimensions 200 x 300 x 80 mm Weight 2 8 kg 3 5 kg 3 5 kg Air Humidit...

Страница 3: ...Adjusting is required only when it is relocated not before daily operation 3 Unpacking and Accessories Carefully unpack the balance and keep the packaging for possible future use when transporting the...

Страница 4: ...ecurely See b 6 Electrical Supply The mains Adapter should only be connected to a 230 V 50 Hz electrical supply Do not connect to any other mains voltages unless another mains voltage is specifically...

Страница 5: ...rements of the following EC directives EC Council Directive on Electromagnetic Compatibility Version 89 336 EC Applied harmonised norms in particular EN 50081 1 EN 50082 1 If a change is made to the a...

Страница 6: ...wer MODE Key for changing display pcs weight or unit weight Also for calling function for storing parameters for setting digits of parameters 10 Key for sampling unit weight with 10 samples 30 Key for...

Страница 7: ...ark Lower Indicates the balance mode is in Unit Weight Display Mode 9 Installation Location KERN EC balances are very robust still it is an precision weighing instrument which requires gentle operatio...

Страница 8: ...ed 10 pieces or 30 pieces on the pan Press 10 key or 30 key corresponding to the sample quantity 3 The display turns blank while lighting small M mark on the left Then a piece display 10 p or 30 p app...

Страница 9: ...for Auto Zero appears To change the parameter at the last digit hit T key As to contents of parameters see next page 3 By hitting Mode the function item will advance successively according to followin...

Страница 10: ...with output option 41 o c 0 No output 1 Constant serial output 2 Constant serial output of stabilised data Output 3 Nil condition 4 Automatic output with a load after stabilisation at the port 5 One o...

Страница 11: ...appears Unit of weight 3 The parameter after unit shows weight unit for adjusting currently linked with the set weight unit in 5 un _ See page 9 Zero setting 4 The display will indicate on 0 Verify t...

Страница 12: ...uld enter the measuring cell Therefore only clean the balance with a dry or damp cloth Do not use solvents as varnished or plastic parts could be damaged A warming up time of a few minutes after switc...

Страница 13: ...changed by relocation or after long time lapse Adjust the balance referring to page 10 Wrong linearity Characteristics have changed or mechanism adjustment has changed by some reason Contact shipper...

Страница 14: ...unit weight of samples is too light for the balance division Countable unit weight is the readability of the balance or over o Err The load exceeds the capacity of the balance The tare is too heavy u...