- 7 -
DAB-SH-e-1610
2) After successful adjustment the balance automatically
returns to weighing mode, „100,000 g“ will be displayed.Take
away adjustment weight.
3) Wait until „0.000 g“ appears.
The
balancecan not
weighing, and
the liner error
more than
±10d.
Three point weight adjustment
Press”
” when the light turn on, release and press
Max load (100g) is flashing
put on the calibration weight (100g F2 is recommended)