CH-BA-defsi-0422
7
8
Maintenance, upkeep, disposal
8.1 Cleaning
Only use a cloth dampened with mild suds and not aggressive
cleaning agents (solvents or similar). Please ensure that fluids
are not able to get into the device and rub off using a clean,
soft cloth. Loose sample residue/powder can be removed
carefully using a brush or hand vacuum cleaner.
8.2 Maintenance,
upkeep
The operating company shall ensure that load-bearing com-
ponents are checked by a competent person at maximum
intervals of one year.
These checks can be necessary at intervals shorter than one
year depending on the conditions of use. This applies, for
example, in the case of particularly frequent use, increased
wear and corrosion.
The device may only be opened by trained service engineers
authorised by KERN.
Disconnect from the mains supply before opening.
8.3 Disposal
The operating company shall dispose of the packaging and
the device in compliance with the valid national or regional law
of the operating location.
9 Troubleshooting
The balance should be switched off for a short time following
an interruption in the programme sequence. It is then neces-
sary to repeat the weighing process from the beginning.
Interruption Possible
cause
•
The balance is not switched on.
Weight display is
not illuminated.
•
The batteries are wrongly inserted, the
batteries are empty
•
No batteries are attached
•
Draught/air movement
•
Table/floor vibrations
The weight dis-
play changes
continually
•
Vibrations on the balance suspension
•
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
•
The balance display is not set to zero
•
Adjustment is no longer correct.
The weighing
result is obviously
incorrect
•
Great fluctuations in temperature.
•
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
Switch the balance off if other error messages should appear
and then switch on again. Contact the manufacturer if the error
message does not disappear.