KERN CDS 100K0.5 Скачать руководство пользователя страница 10

4  Basic Safety Precautions 

4.1  Pay attention to the instructions in the Operation Manual 

 

Carefully read this operation manual before setup and commissioning, even 
if you are already familiar with KERN balances. 

Versions in other languages are non-binding translations. 
The only binding version is the original document in German.

 

4.2  Personnel training 

The appliance may only be operated and maintained by trained personnel. 
 
 
 

5  Transport and storage 

5.1  Testing upon acceptance 

When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking 
for possible visible damage. 
In case of visible damage have the damage verified by the messenger’s signature.  Do not alter goods or 
packaging and do not remove any parts of the delivery. Report the damage immediately (within 24 hours) in 
writing to the parcel service. 

 

 

5.2  Packaging / return transport 

 

 

Keep all parts of the original packaging for a possibly required return. 

 

Only use original packaging for returning. 

 

Prior to dispatch disconnect all cables and remove loose/mobile parts.  

 

Reattach possibly supplied transport securing devices.  

 

Secure all parts such as the glass wind screen, the weighing platform, power 
unit etc. against shifting and damage. 

 

 
 

CKE/CDS-BA-e-1526 

10 

Содержание CDS 100K0.5

Страница 1: ...H Ziegelei 1 D 72336 Balingen email info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating Manual Counting balances KERN CKE CDS Version 2 6 01 2015 GB C...

Страница 2: ...lacing 11 6 3 Mains connection 11 6 4 Battery operation CKE 12 6 5 Connection of peripheral devices 12 6 6 Initial Commissioning 12 6 7 Adjustment 13 6 8 Adjusting see chapter 7 2 1 13 6 9 Verificatio...

Страница 3: ...tare key and press the ON OFF key Let go of the tare key Using the automatic reference optimization OPT the counting accuracy is automatically increased when applying up to 100 items Factory settings...

Страница 4: ...3 0 3 6 kg 2 4 5 6 kg Recommended adjustment weight F1 not added 300 g 2 kg 2 kg 1 kg 5 kg Appropriate for verification max 80 rel non condensing Stabilization time typical 3 sec Allowable ambient te...

Страница 5: ...7 8 kg 5 10 15 16 kg 5 10 15 16 kg Recommended adjustment weight F1 not added 5 kg 2 kg 10 kg 5 kg 10 kg 5 kg Appropriate for verification max 80 rel non condensing Stabilization time typical 3 sec A...

Страница 6: ...kg 20 30 50 60 kg 20 30 50 60 kg Recommended adjustment weight F1 not added 20 kg 10 kg 50 kg 50 kg Appropriate for verification max 80 rel non condensing Stabilization time typical 3 sec Allowable a...

Страница 7: ...t point kg 1 2 4 2 5 10 15 2 5 10 15 16 10 15 20 30 10 15 20 30 Recommended adjustment weight F1 not added 2 kg 2 kg 10 kg 5 kg 10 kg 5 kg 20 kg 10 kg 20 kg 10 kg Appropriate for verification max 80 r...

Страница 8: ...36 20 30 50 60 20 50 100 Recommended adjustment weight F1 not added 20 kg 10 kg 50 kg 50 kg 50 kg Appropriate for verification max 80 rel non condensing Stabilization time typical 3 sec Allowable amb...

Страница 9: ...in explosive environment The serial version is not explosion protected The structure of the balance may not be modified This may lead to incorrect weighing results safety related faults and destructio...

Страница 10: ...ing immediately and the appliance itself when unpacking for possible visible damage In case of visible damage have the damage verified by the messenger s signature Do not alter goods or packaging and...

Страница 11: ...n of air humidity on the appliance may occur if a cold appliance is taken to a considerably warmer environment In this case acclimatize the disconnected appliance for ca 2 hours at room temperature Av...

Страница 12: ...tical value for operational safety this will be indicated in the display with the BATT LOW information 6 5 Connection of peripheral devices Before connecting or disconnecting of additional devices pri...

Страница 13: ...warming up time of ca 15 minutes is recommended for stabilization 6 9 Verification General introduction According to EU directive 90 384 EEC balances must be officially verified if they are used as f...

Страница 14: ...nce using 10 parts Generate reference using 20 parts In mode menu YES function Switchover pcs g In menu Mode function Print out weighing result Display icon Significance OVERLOAD Overload Weighing ran...

Страница 15: ...7 2 Operation 7 2 1 Adjusting 7 2 2 Speed 7 2 3 Auto Off CKE CDS BA e 1526 15...

Страница 16: ...CKE CDS BA e 1526 16...

Страница 17: ...o select the menu point Backlight auto and confirm by pressing the YES button The background illumination will be switched off automatically 10 sec after having reached a stable weighing value 7 2 5 S...

Страница 18: ...ground Baud rate The MODE key is used to set the Baud rate used for the transfer of measured values In the following example the baud rate is set to 4800 baud Setting Baud rate KERN CKE Display 1 Swit...

Страница 19: ...f When AUTOPRINT PC is active the current weighing values will be sent continuously via the RS 232 data interface Data output and remote control commands Remote control commands transferred as ASCII c...

Страница 20: ...e to their size or shape can be weighed with the help of the underfloor weighing facility Proceed as follows Switch off balance Turn over the balance and in doing so take care that the weighing plate...

Страница 21: ...case of an error in the program process briefly turn off the balance and disconnect from power supply The weighing process must then be restarted from the beginning Help Fault Possible cause The displ...

Страница 22: ...duit auquel se rapporte la pr sente d claration est conforme aux normes cit es ci apr s I Dichiarazione di conformit Dichiariamo con ci che il prodotto al quale la presente dichiarazione si riferisce...

Отзывы: