![KERN 770-12 Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/kern/770-12/770-12_operating-instructions-manual_1957145055.webp)
122
770/GS/GJ-BA-defsi-0023
Control Commands with lower-case Letters
All functions that can be selected by pressing the respective keys on the balance can
also be activated by commands.
ESC
f
0
-
CR
LF
Function key F
ESC
f
1
-
CR
LF
Function key CAL
ESC
s
3
-
CR
LF
Clear Function CF
ESC
x
0
-
CR
LF
Perform sensitivity test*
ESC
x
1
-
CR
LF
Output balance model
ESC
x
2
-
CR
LF
Output serial number
Each control command with the lower-case letters f, s and x must be terminated
by an underline (ASCII = 95).
* = only for balances with a built-in automatic adjusting function