![KERN 770-12 Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/kern/770-12/770-12_operating-instructions-manual_1957145051.webp)
118
770/GS/GJ-BA-defsi-0023
Special Codes
Special codes are output only if the balance operating menu code 6 1 1, 6 1 4 or 6 1 5 is
set (see the section entitled “Data Output Parameters“).
Special status-dependent codes
1 2 3 4 5 6 7 8 9 10
11
12
13
14
15
16
* * * * * * A B * * * * * * CR
LF
The following status codes are output for “A B“:
* * : Tare
H * : Overload
C * : Adjust*)
L * : Underload
- - : All numerals indicated in stable readout
Special error-dependent codes:
1 2 3 4 5 6 7 8 9 10
11
12
13
14
15
16
* * * E R R * X Y Z * * * * CR
LF
X
= *, 0, 1, 2 as a one-place error code
Y Z = two-place error index code
* = space
*) = For balances with a built-in automatic adjusting function, the displayed status
code “C“ will also be output when a print command is received.