440-BA-e-1744
6
KERN
440-47N
440-49N
440-49A
Readability (d)
0.1 g
0.1 g
0.1 g
Weighing range (max)
2000 g
4000 g
6000 g
Taring range (subtractive)
2000 g
4000 g
6000 g
Reproducibility
0.1 g
0.1 g
0.1 g
Linearity
± 0.2 g
± 0.3 g
± 0.3 g
Minimum unit weight at piece
count
0.2 g
0.2 g
0.2 g
Warm-up time
30 minutes
2 hours
2 hours
Reference unit weights at piece
count
5, 10, 25, 50
Weighing Units
Details
„Weighing units“
chapter 9.3
Recommended adjustment
weight, not added (class)
2000 g (M1)
4000 g (F2)
5000 g (F2)
Stabilization time (typical)
3 sec.
Operating temperature
+ 5° C .... + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
165 x 230 x 80
Weighing plate mm
130 x 130
150 x 170
150 x 170
Weight kg (net)
1.0
1.0
1.0
Underfloor weighing
Standard
Flush-mounted hook
Option
Содержание 440-21A
Страница 75: ......