KERN 1500-2M Скачать руководство пользователя страница 13

EW-BA-defsi-0620

28

Baud rate 

7.  b.L. 

1*. 
2
3

1200 Baud 
2400 Baud 
4800 Baud 

Unit commutation 

8.  Set. 

1* 
2
3
4

g
g/ct
not documented 
not documented 

Not documented 

0.  GLP 

1  

Always use this setting 

Not documented 

A.  PrF. 

Always use this setting 

11  Important information 

  These  electronic  balances  are  precision  instruments.  It  is  designed  for  use  in  a 

self-controlled  electromagnetic  environment.  That  means  that  in  such  an 
environment radio equipment, such as mobile phones, must not be used in direct 
vicinity.  In  case  strong  electromagnetic  fields  should  arise,  deviations  of  the 
indication are possible. Avoid disturbing environmental conditions such as currents 
and vibrations as well as a rapid change of the ambient temperature. If necessary 
readjust the balance after a change of temperature. 

  Avoid  high  humidity,  vapours  and  dust  as  the  balance  is  not  hermetically  closed. 

Do  not  let  the  balance  come  into  direct  contact  with  liquids  as  these  could  enter 
the measuring cell. Therefore only clean the balance with a dry or damp cloth. Do 
not use solvents as varnished or plastic parts could be damaged. 

  A warming up time of a few minutes after switching the balance on will stabilise the 

measuring  values.  Place  the  object  carefully  on  the  weighing  pan.  Avoid  any 
constant  load  on  the  weighing  pan  even  when  out  of  use.  Avoid  shocks,  thrusts 
and extreme impairments by all means. Remove spillages immediately. 

  When the programme cycle is disturbed switch the balance off and then on again. 

Start the weighing operation again. 

  Never use the balance in rooms where there is a danger of explosion. The serial 

version is not explosion-proof. 

  Opening of the balance causes invalidity of warranty. 
  Check the balance daily with the supplied check weight. 

  If necessary adjust the balance. 

  The warranty is nullified if the balance is opened. 

Содержание 1500-2M

Страница 1: ......

Страница 2: ...Recommended adjusting weight not included class 1 x 100g 1 x 20g F1 500 g F2 1 x 1 kg 1 x 200 g F1 Stabilisation time 2 sec Weighing plate stainless steel 110 mm 140 mm 140 mm Balance dimensions 180 x...

Страница 3: ...weight not included class 1 x 2 kg 1 x 1kg E2 5 kg F2 Stabilisation time 2 sec Weighing plate stainless steel 140 mm Balance dimensions 180 x 230 x 85 mm Mains adapter 230 V 50 60 Hz balance 9 V DC 4...

Страница 4: ...Weighing pan support External mains adapter Operating instructions 3 Setting up the balance Place the pan base packed with the weighing pan on the balance Fix it on the shaft by driving the knurled nu...

Страница 5: ...si 0620 20 4 External view Key functions ON OFF ON OFF key for lighting display Print Print key or data transmission key F Function key to call various functions TARE Zeroing Taring key parameter sele...

Страница 6: ...y 5 minutes after turning it on 6 Power supply Plug balance into a socket 230 V 50 Hz Do not connect it when there should be another mains voltage Socket must be grounded as per VDE prescription Inspe...

Страница 7: ...t the electromagnetic compatibility Version 90 384 EWG Version 89 336 EWG Applied harmonised norms in particular EN 45501 EN 50081 1 EN 50082 1 This declaration is only valid with the certificate of c...

Страница 8: ...ls KERN EW Blocking Releasing of the adjusting function Remark The adjusting is only possible when not being blocked by the adjusting switch Switch to the right Adjusting function is blocked After suc...

Страница 9: ...nutes is necessary Press F and hold pressed until CAL appears then release the key Press Tare and hold pressed Immediately press F briefly and then also release TARE Display on 0 briefly flashes Do no...

Страница 10: ...us the power display in the balance display no longer is visible It is recommended to turn the mains adapter off after having terminated to work with the balance Bar Graph display The weighing range o...

Страница 11: ...the cup has been stored now internally Put the load into the cup and read off the weighing value The tare process can be repeated as often as wanted for example to weigh in several components as a mix...

Страница 12: ...ighing display 1 b0G 0 1 Balance display is turned off Balance display is turned on Auto Zero 3 A 0 0 1 deactivated activated Automatic shut down Accu only 4 A P 0 1 deactivated Balance is shut down a...

Страница 13: ...vapours and dust as the balance is not hermetically closed Do not let the balance come into direct contact with liquids as these could enter the measuring cell Therefore only clean the balance with a...

Страница 14: ...g range is exceeded b Err Electronic error caused by electrostatic field Check environment If necessary choose another location u Err The weighing pan does not lie on the pan support 1 Err Wrong adjus...

Отзывы: