background image

11 

 

OZL-45R-BA-e-1510 

5.4  Dioptre adjustment and focussing 

 
A  special  feature  of  stereo  microscopes  is  that  they  are  fitted  with  an  optical  unit 
which has a relatively high depth of field. In order to be able get the most benefit from 
this feature, each user must synchronise the focussing mechanisms for themselves. 
 
The steps to do this are described in the following section. 
 
 
1.  Place the object to be observed on the working surface under the objective. 

 

2.  Put both dioptre adjustment rings into the starting position of 0. 
 
3.  Use the zoom control dials to set the smallest possible magnification. 
 
4.  Look through the right eyepiece with the right eye and bring the object into focus 

by using the focus control dials. 

 
5.  Now set the largest possible zoom factor. 
 
6.  Once  again,  still  only  looking  through  the  right  eyepiece,  bring  the  object  into 

focus 

 
7.  Then set the smallest possible zoom factor again. 
 
8.  If the object then does not appear to be in focus, adjust the focus on the dioptre 

adjustment ring of the right eyepiece. 

 
9.  In order to get the highest level of accuracy when adjusting the focus, you should 

repeat steps 5-8. 

 
10. Afterwards set back to the smallest possible zoom factor. 
 
11. Then look through the left eyepiece with the left eye and use the lefthand dioptre 

adjustment ring to also adjust the optimum focus of the object. 

 
12. In this way, the object being observed will be in focus at any zoom setting. 
 
 

 

Содержание OZL 456

Страница 1: ...RN Sohn GmbH Ziegelei 1 D 72336 Balingen E mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Stereo zoom microscope KERN OZL 45R OZL 456...

Страница 2: ......

Страница 3: ...7 4 Assembly 8 5 Operation and functionality 10 5 1 Getting started 10 5 2 Adjusting the interpupillary distance 10 5 3 Adjusting the magnification 10 5 4 Dioptre adjustment and focussing 11 5 5 Adjus...

Страница 4: ...l parts in the base of the device must not be tampered with as in this area there is an additional risk of triggering an electric shock 1 2 Notes on the electrical system Before connecting to a mains...

Страница 5: ...gid smooth horizontal surface When the microscope is not being used you should fit the objective cap and cover the microscope with the enclosed dust protective cover If the eyepieces are being stored...

Страница 6: ...s these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However organ...

Страница 7: ...OZL 45R BA e 1510 6 2 Nomenclature...

Страница 8: ...e angled at 45 Interpupillary distance 55 75 mm Dioptre adjustment On both sides Packaging dimensions WxDxH 345x320x470 mm Standard configuration Model KERN Tube Eyepiece Field of view mm Objective Zo...

Страница 9: ...ctors so that you can then fit the eyepieces When doing this please be particularly careful that you do not touch the optical lenses with your fingers and that no dust enters the apertures You should...

Страница 10: ...9 OZL 45R BA e 1510 Assembled stereo zoom microscope...

Страница 11: ...ieces must be re adjusted While you are looking through the eyepieces use one hand to hold the righthand or lefthand prism housing firmly By rotating outwards or inwards you can either increase or red...

Страница 12: ...ough the right eyepiece with the right eye and bring the object into focus by using the focus control dials 5 Now set the largest possible zoom factor 6 Once again still only looking through the right...

Страница 13: ...the other wheel Depending on the direction of the turn the torque will be increased or decreased On one hand this function can help to make it easier to adjust the focus and on the other hand it can p...

Страница 14: ...ied by the glasses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved W...

Страница 15: ...reflected light is on the right rear corner The reflected light is provided by an LED ring lighting unit which is fitted directly under the objective and is connected to the objective Furthermore you...

Страница 16: ...is placed next to in front of or behind the microscope For halogen lighting units the light source is in the housing of the unit and is emitted through one or several optical fibre cables For LED unit...

Страница 17: ...75x 62 5x 28 125x 187 5x 37 5x 255x Field of view mm 15 8 2 4 31 5 4 8 10 5 1 6 7 9 1 2 Working distance 113 mm 220 mm 50 mm 35 mm 7 Features Model outfit Kern model Order number OZL 456 Eyepieces HWF...

Страница 18: ...sufficient The wrong bulb has been used The input voltage is too low Optical unit Problem Possible causes You can see two images The interpupillary distance is not set correctly The magnifications of...

Страница 19: ...y materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...

Отзывы: