25 

 

OBD-1-BA-e-1510 

Problem 

Possible causes 

Blurred details 
 
Bad image 
 
Bad contrast 
 
Vignetted field of view 

The aperture diaphragm is not opened wide 
enough 

The condenser is too low 

The objective does not belong to this 
microscope 

The front lens of the objective is dirty 

An immersion object has been used without 
immersion oil 

The immersion oil contains air bubbles 

The condenser is not correctly centred 

The recommended immersion oil has not 
been used 

Dirt / dust on the objective 

Dirt /dust on the front lens of the condenser 

Dirt or dust in the field of view  

Dirt / dust on the eyepieces 

Dirt / dust on the front lens of the condenser 

Dirt / dust on the object 

One side of the image is blurred 

The stage was not correctly fitted 

The objective is not positioned correctly on 
the beam path 

The nosepiece is not correctly fitted 

The upper side of the object is facing down 

The image flickers 

The nosepiece is not correctly fitted 

The objective is not positioned correctly on  
the beam path 
The condenser has not been correctly  
centred 

The coarse adjustment knob is difficult to 
turn 

The rotational resistance brake is too  
tight 
The angle table is blocked by a  
solid body 

The stage moves down on its own 
The fine adjustment knob moves on its 
own 

The rotational resistance brake is not tight 
enough 

When you move the table, the image 
becomes blurred 

The stage was not correctly fitted 

 

 

Содержание OBD 127

Страница 1: ...gelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Transmitted light laboratory microscope digital KERN OBD 1 O...

Страница 2: ......

Страница 3: ...4 4 Colour filter 10 4 5 Condenser 11 5 Operation 11 5 1 Getting started 11 5 2 Pre focussing 12 5 3 Adjusting the interpupillary distance 13 5 4 Dioptre adjustment 13 5 5 Adjusting the magnification...

Страница 4: ...n additional risk of triggering an electric shock 1 2 Notes on the electrical system Before connecting to a mains power supply you must make sure that you are using the correct input voltage The infor...

Страница 5: ...ch are too high or too low vibrations dust or a high level of humidity The ideal temperature range is between 0 and 40 C and a relative humidity of 85 should not be exceeded The device should always b...

Страница 6: ...these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However organi...

Страница 7: ...OBD 1 BA e 1510 6 2 Nomenclature...

Страница 8: ...7 OBD 1 BA e 1510 Rear view...

Страница 9: ...ty Trinocular 6V 20W Halogen Transmitting OBD 128 Infinity Trinocular 6V 20W Halogen Transmitting Eyepieces WF 10x 20 mm Objectives 4x 10x 40x 100x Product dimensions 395x200x430 mm Packaging dimensio...

Страница 10: ...9 OBD 1 BA e 1510...

Страница 11: ...on appears When the objectives need to be dismounted you should always make sure that you do not touch the lenses with your bare fingers and that no dust enters the apertures For objectives which are...

Страница 12: ...irst step is to establish a power connection using the mains plug You should first adjust the dimmer to a low level so that when you look through the eyepiece for the first time your eyes are not imme...

Страница 13: ...soon as an image is recognisable no matter how sharp then you should only adjust the focus using the fine adjustment knob Adjusting the torque of the coarse and fine adjustment knob Next to the left a...

Страница 14: ...cular image then the interpupillary distance is set correctly 5 4 Dioptre adjustment The eye strengths of each eye of the microscope user can often be slightly different which in daily life has no con...

Страница 15: ...objectives you should use the silver ring above the objectives see illustration When rotating the nosepiece you must always make sure that the objective which is about to be positioned in the beam pat...

Страница 16: ...is is regarded as Fixed Koehler since the condenser is pre centred When adjusting the Koehler lighting for the first time you must first select the lowest possible objective magnification so that you...

Страница 17: ...illustration on the right The diameter of the aperture diaphragm which is then visible should make up approximately 2 3 of the pupil diameter If the eyepiece should be removed for checking then please...

Страница 18: ...must establish a connection between the USB port of the camera and an USB port of a PC The needed cable is included in the scope of delivery Furthermore there is also a disk that enables you to instal...

Страница 19: ...sses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved We would theref...

Страница 20: ...Lower the specimen stage and position the 100x objective in the beam path 3 Bring the specimen stage or object slide to the objective very slowly until there is slight contact 4 Observe the object The...

Страница 21: ...is on the underside of the device It can be opened by undoing the screws on the holder see illustration We recommend that here you should also test again to check that heat is no longer being produce...

Страница 22: ...polarisation unit consists of a polariser and an analyser Both consist of a round glass plate with an external holder ring To mount the analyser you have to remove the microscope head at first Then t...

Страница 23: ...y out dark field applications 1 A special dark field condenser can be used in place of the standard condenser This as a paraboloid construction and also meets the requirements of professional applicat...

Страница 24: ...ass without being affected by the PH slider The PH objective has a PH ring in its lens system just like the PH slider Due to their size on one hand and their positioning on the other these two rings m...

Страница 25: ...y centred The condenser is too low The field of view is dark or is not correctly illuminated The objective is not positioned correctly on the beam path The selector switch for the beam path is between...

Страница 26: ...Dirt dust on the front lens of the condenser Dirt dust on the object One side of the image is blurred The stage was not correctly fitted The objective is not positioned correctly on the beam path The...

Страница 27: ...materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...