5.6  Adjusting the Koehler illumination 

 
To  make  sure  that  perfect  image  results  are  achieved  during  microscopic 
observation, it is important that the direction of light of the microscope is optimised. If, 
as with the devices in the KERN OBD-1 series, the lighting can be set in accordance 
with  Koehler,  the  result  is  homogenous  illumination  of  the  slide  and  avoidance  of 
disruptive stray light.  
 
The necessary control elements for this are: 

 

  Height-adjustable and centre-adjustable condenser with aperture diaphragm 

  Field diaphragm 

 
Note:  The  condenser  of  the  microscopes  of  the  series  OBD-1  is  not  centre-
adjustable. Therefore it is not a full Koehler illumination. However this is regarded as 
Fixed-Koehler, since the condenser is pre-centred. 
 
When adjusting the Koehler lighting for the first time, you must first select the lowest 
possible objective magnification, so that you can carry out the following steps. 
 
 
 
 
1.  Use  the  condenser  focus  wheel  to  position  the  condenser 

directly  below  the  specimen  stage.  Switch  on  the  lighting 
and  use  the  coarse  and  fine  adjustment  knob  to  bring  the 
slide  with  the  cover  glass  positioned  facing  upwards  into 
focus. 

 
 
 
 
 
 
2.  Close  the  field  diaphragm  completely  using  its  adjusting 

ring.  When  you  look  in  the  microscope  a  blurred  image 
appears in the aperture. 

 
 
 
 
 
 
3.  Adjust the height of the condenser until the image from the 

field  diaphragm  appears  clearly  in  the  field  of  view.  For 
some  microscopes  there  is  a  risk  that  you  will  lift  the 
condenser up so high that it collides with the object holder. 
Therefore care is needed when doing this. 

 
 

 

Содержание OBD 127

Страница 1: ...gelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Transmitted light laboratory microscope digital KERN OBD 1 O...

Страница 2: ......

Страница 3: ...4 4 Colour filter 10 4 5 Condenser 11 5 Operation 11 5 1 Getting started 11 5 2 Pre focussing 12 5 3 Adjusting the interpupillary distance 13 5 4 Dioptre adjustment 13 5 5 Adjusting the magnification...

Страница 4: ...n additional risk of triggering an electric shock 1 2 Notes on the electrical system Before connecting to a mains power supply you must make sure that you are using the correct input voltage The infor...

Страница 5: ...ch are too high or too low vibrations dust or a high level of humidity The ideal temperature range is between 0 and 40 C and a relative humidity of 85 should not be exceeded The device should always b...

Страница 6: ...these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However organi...

Страница 7: ...OBD 1 BA e 1510 6 2 Nomenclature...

Страница 8: ...7 OBD 1 BA e 1510 Rear view...

Страница 9: ...ty Trinocular 6V 20W Halogen Transmitting OBD 128 Infinity Trinocular 6V 20W Halogen Transmitting Eyepieces WF 10x 20 mm Objectives 4x 10x 40x 100x Product dimensions 395x200x430 mm Packaging dimensio...

Страница 10: ...9 OBD 1 BA e 1510...

Страница 11: ...on appears When the objectives need to be dismounted you should always make sure that you do not touch the lenses with your bare fingers and that no dust enters the apertures For objectives which are...

Страница 12: ...irst step is to establish a power connection using the mains plug You should first adjust the dimmer to a low level so that when you look through the eyepiece for the first time your eyes are not imme...

Страница 13: ...soon as an image is recognisable no matter how sharp then you should only adjust the focus using the fine adjustment knob Adjusting the torque of the coarse and fine adjustment knob Next to the left a...

Страница 14: ...cular image then the interpupillary distance is set correctly 5 4 Dioptre adjustment The eye strengths of each eye of the microscope user can often be slightly different which in daily life has no con...

Страница 15: ...objectives you should use the silver ring above the objectives see illustration When rotating the nosepiece you must always make sure that the objective which is about to be positioned in the beam pat...

Страница 16: ...is is regarded as Fixed Koehler since the condenser is pre centred When adjusting the Koehler lighting for the first time you must first select the lowest possible objective magnification so that you...

Страница 17: ...illustration on the right The diameter of the aperture diaphragm which is then visible should make up approximately 2 3 of the pupil diameter If the eyepiece should be removed for checking then please...

Страница 18: ...must establish a connection between the USB port of the camera and an USB port of a PC The needed cable is included in the scope of delivery Furthermore there is also a disk that enables you to instal...

Страница 19: ...sses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved We would theref...

Страница 20: ...Lower the specimen stage and position the 100x objective in the beam path 3 Bring the specimen stage or object slide to the objective very slowly until there is slight contact 4 Observe the object The...

Страница 21: ...is on the underside of the device It can be opened by undoing the screws on the holder see illustration We recommend that here you should also test again to check that heat is no longer being produce...

Страница 22: ...polarisation unit consists of a polariser and an analyser Both consist of a round glass plate with an external holder ring To mount the analyser you have to remove the microscope head at first Then t...

Страница 23: ...y out dark field applications 1 A special dark field condenser can be used in place of the standard condenser This as a paraboloid construction and also meets the requirements of professional applicat...

Страница 24: ...ass without being affected by the PH slider The PH objective has a PH ring in its lens system just like the PH slider Due to their size on one hand and their positioning on the other these two rings m...

Страница 25: ...y centred The condenser is too low The field of view is dark or is not correctly illuminated The objective is not positioned correctly on the beam path The selector switch for the beam path is between...

Страница 26: ...Dirt dust on the front lens of the condenser Dirt dust on the object One side of the image is blurred The stage was not correctly fitted The objective is not positioned correctly on the beam path The...

Страница 27: ...materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...

Отзывы: