SYSTem:SET COMMAND
SYST:SET
Syntax
:
Short Form: SYST:SET {SA0 | SA1 | CM0 | CM1 | SI0 |}
LONG Form: SYSTem:SET {SA0 | SA1 | CM0 | CM1 | SI0 |}
Description:
Allows the unit to behave as a standard Kepco MST 488-27 Controller (programmer).
Sending
SYST:SET SA0 disables the default secondary address mode of the MST 488HT controller and
causes the unit to revert to the command syntax of the non-HT suffix MST 488-27 Controller (Pro-
grammer).
To return the unit to secondary addressing send SYST:SET SA1
After sending SYST:SET SA0, SCPI commands and queries behave as documented in Appendices A
and B of MST 488-27 Programmer Operator Manual except as noted below. The MST 488-27 Pro-
grammer Operator Manual can be downloaded from the Kepco website at
https://www.kepcopower.com/support/opmanls.htm#mst
After sending SA0 the following exceptions apply:
•
Status reporting bits for STAT:OPER registers and STAT QUES registers follow the assignments
listed below in Tables B-3 and B-4 respectively.
•
The unit does not support SYST:SET SI1 which enables the STAT:INST register status reporting
capability of the non-HT MST 488-27 unit.
•
Some MST 488-27HT commands are still supported.
— VOLT:PROT to allow for over voltage protection setting
— OUTP:PROT:DEL current mode timing delay
— OUTP:PROT:CLE protection reset
— OUTP:REL output relay enable and disable
— OUTP:REL:POL? allows user to determine if a positive or negative output is enabled.
TABLE B-3. OPERATION CONDITION REGISTER, OPERATION ENABLE REGISTER,
AND OPERATION EVENT REGISTER BITS
CONDITION
STC
NU
CC
NU
CV
NU
NU
WTG
NU
NU
NU
NU
CAL
BIT
12
11
10
9
8
7
6
5
4
3
2
1
0
VALUE
4096
2048
1024
512
256
128
64
32
16
8
4
2
1
CAL CALIBRATION
CC
CONSTANT CURRENT
CV
CONSTANT VOLTAGE
NU
NOT USED
STC LAST STEP COMPLETE
WTG WAITING FOR TRIGGER
TABLE B-4. QUESTIONABLE EVENT REGISTER, QUESTIONABLE CONDITION REGISTER
AND QUESTIONABLE CONDITION ENABLE REGISTER BITS
CONDITION
NU
UNR
RI
NU
NU
NU
NU
OT
NU
NU
OC
OV
BIT
11-15
10
9
8
7
6
5
4
3
2
1
0
VALUE
1024
512
256
128
64
32
16
8
4
2
1
NU
NOT USED
OC
OVERCURRENT PROTECT
OT
OVERTEMPERATURE
OV
OVERVOLTAGE PROTECT
RI
REMOTE INHIBIT
UNR OUTPUT UNREGULATED
Содержание MST 488-27HT
Страница 2: ......
Страница 8: ...iv MST488 27 SVC 051120 FIGURE 1 1 MST488 27 POWER MODULE CONTROLLER ...
Страница 11: ...MST 488 27HT 051120 1 3 FIGURE 1 2 REMOTELY CONTROLLED POWER SUPPLY CONFIGURATIONS USING KEPCO PRODUCTS ...
Страница 12: ...1 4 MST 488 27HT 051120 FIGURE 1 3 MST 488 27HT CONTROLLER OUTLINE DRAWING ...
Страница 14: ......
Страница 20: ...2 6 MST 488 27 051120 FIGURE 2 2 FRONT AND REAR PANELS OF MST 488 27HT POWER MODULE CONTROLLER ...
Страница 22: ...2 8 MST 488 27 051120 ...
Страница 28: ......
Страница 34: ......