Kentech Instruments Pulse Chopper System Скачать руководство пользователя страница 7

7

Kentech Instruments Ltd., Unit 9, Hall Farm Workshops, South Moreton, Didcot, Oxon, OX11 9AG, U.K.

10th. July 1999

The following is a procedure we have used for timing the cells.

1

Set up the system and align as under static alignment.

2

It is much easier to look for transmission than extinction when one is using each channel of the pulser
independently. For timing tests it is better to rotate the output polariser, so that it is perpendicular
to that at the input.

3

Disconnect the DC cabling and connect one channel of the pulser to one cell. Note the timing of the
edge of the transmission curve with an oscilloscope.

4

Disconnect the pulser and in turn work through the other three channels, one at a time. Note that the
pulser must be connected to the cell that it will end up being used with. You cannot mix the channels
and cells as the cells require slightly differently timed drive pulses to achieve the same output timing.

5

Once all the timings are noted use one in approximately the centre of the range as a reference and
adjust all the others so that they are the same; see section 5.3.

Note the unit supplied is set up so that the timing is correct when the light enters from the right in figure 3 and

the pulser channels are connected as shown.

5.3

ADJUSTING THE TIMING OF THE CHANNELS.

The timing to each channel is adjusted by means of a set of four potentiometers on the trigger board. These are

a little hard to access. Care must be taken if operating the unit without the covers on the pulser. We strongly recommend
that the pulser is turned off and the mains disconnected before each adjustment is made, unfortunately continuous
adjustment whilst observing the effect is not possible with the unit turned off. It is likely that only small changes to the
timing will be necessary.

To change the timing:-

1

Turn off and remove the power from the unit

2

Remove the eight small cross point screws in the two sides of the unit.

1

2

3

4

Figure  3

The head showing the connections and direction of light propagation

Содержание Pulse Chopper System

Страница 1: ...shops South Moreton Didcot Oxon OX11 9AG U K 10th July 1999 Unit 9 Hall Farm Workshops South Moreton Didcot Oxon OX11 9AG U K Kentech Instruments Ltd Pulse Chopper System J99 10th July 1999 PLEASE REA...

Страница 2: ...capacity is small careless use could result in electric shock It is assumed that this highly specialised equipment will only be used by qualified personnel The manufacturers and suppliers accept no re...

Страница 3: ...ING THE TIMING OF THE CHANNELS 7 6 0 MOUNTING THE CELLS 9 6 1 PHYSICAL MOUNTING 9 6 2 ELECTRICAL MOUNTING 9 6 3 DC CONNECTIONS 9 Figure Captions Figure 1 The front panel 5 Figure 2 A plot of DC transm...

Страница 4: ...ontrol controls the pulse voltage in approximately 20 volt steps from 210 volts to 470 volts All four channels are adjusted together This is effectively the voltage into a 50 load By mismatching the e...

Страница 5: ...amplitude polarised at 90 to each other These travel at slightly different speeds in the excited medium and emerge with a relative phase shift of 2 In the unexcited medium there is no relative change...

Страница 6: ...s easily obtained from the transmission obtained when the voltage is removed to the transmission when the voltage is applied Note that some power supplies will not discharge the cells when they are tu...

Страница 7: ...ming 5 Once all the timings are noted use one in approximately the centre of the range as a reference and adjust all the others so that they are the same see section 5 3 Note the unit supplied is set...

Страница 8: ...ot Oxon OX11 9AG U K 10th July 1999 Figure 4 Locating the timing adjusters View with bottom cover removed There are four white adjusters The adjuster at the top of thispicture closesttothe bottom pane...

Страница 9: ...ght improve the risetime or not The further apart the cells are mounted the greater the timing for the second cell will have to be adjusted as it will have to be driven slightly later It is possible t...

Страница 10: ...s Ltd Unit 9 Hall Farm Workshops South Moreton Didcot Oxon OX11 9AG U K 10th July 1999 1 11 12 13 14 15 16 17 18 19 2 3 4 5 6 7 8 9 10 Figure 5 Output pulse shape one of four channels 100voltsperdivis...

Страница 11: ...Hall Farm Workshops South Moreton Didcot Oxon OX11 9AG U K 10th July 1999 Figure 6 The head connected for testing Figure 7 Gating off and looking for good extinction Tested with a Kentech Gated Optica...

Страница 12: ...uth Moreton Didcot Oxon OX11 9AG U K 10th July 1999 Figure 8 Testing for long term extinction Showing that reasonable extinction persists to 30ns Figure 9 The gate off fall time Measured with a Kentec...

Страница 13: ...13 Kentech Instruments Ltd Unit 9 Hall Farm Workshops South Moreton Didcot Oxon OX11 9AG U K 10th July 1999 Figure 10 The gate off fall time is measured at 728ps...

Отзывы: