Kent Scientific Rovent Jr. Скачать руководство пользователя страница 7

4

Menus

The Main Menu is the non-touch gateway to all features and setting choices for 

configuring RoVent® Jr., useful when a touch screen is not a feasible option, such as for 

an operator with damp hands. Access the Main Menu through the Touch Screen Menu or 

from the Run Screen by holding Dial A.

Main Menu

Body wgt

Vent

Modules

Config

Power off

Help

Set body weight

Turn Dial B to choose the weight

of your animal

Hold Dial A for Help.

30g

Nominal RR:  133bpm

Nominal Tidal Vol: 0.200mL

Nominal Minute Vol:  26mL/min

Main Menu

(Screen buttons may vary based on the modules installed.)

Alarms

RoVent® Jr. system alarms alert you to certain conditions that might require attention 

while RoVent® Jr. is running. 

There are two levels of alarms:

Level

Warning

Based On

Banner

Sound (If On)

Configurable

Serious

Parameters or other 

operating conditions

Parameters or other 

operating conditions

Yellow

Red

1-beep pattern

3-beep pattern

Yes

Yes except for 

system alarms

To clear a sounding alarm:

A triggered alarm produces a colored banner at the bottom of the screen describing the 

cause of the alarm. The banner color corresponds to the severity of the alarm.

  To silence the alarm:

 Touch:

   

1)   the bottom half of the Run Screen.

   

2)  “Stop Alarm”.

Содержание Rovent Jr.

Страница 1: ...RoVent Jr Automatic Ventilator Volume Pressure Controlled User s Guide...

Страница 2: ...21 Kent Scientific Corporation Documentation Version 2 00 for On Unit Help Version 4 01 and Firmware Revision 1 01 00 36A April 2021 RoVent Jr Automatic Ventilator Volume Pressure Controlled User s Gu...

Страница 3: ...2 Touchscreen 2 Control 2 Display Options 3 Menus 4 Alarms 4 RoVent Jr Automatic Ventilator 5 Set Up Hardware 5 Ventilating an Animal 7 Configure Settings 8 Troubleshooting 9 General Information 11 Co...

Страница 4: ...orized for use in human applications System Components Front Panel Back Panel Exhaust Air Inlet SERIAL MADE IN U S A 1116 Litchfield Street Torrington CT 06790 Tel 860 626 1172 Toll Free 888 572 8887...

Страница 5: ...ch the bottom half of the Run Screen to open the pop up Display Menu and quickly scroll through charts view other data or to silence Alarms Run Screen Power O General Setup Help Body Weight Start Vent...

Страница 6: ...en Hold down See Main Menu on page 4 Display Options Customizing the Run Screen Display To configure Parameters 1 Touch 1 the upper half of the Run Screen for the Touch Screen Menu 2 General setup 3 C...

Страница 7: ...in Main Menu Screen buttons may vary based on the modules installed Alarms RoVent Jr system alarms alert you to certain conditions that might require attention while RoVent Jr is running There are two...

Страница 8: ...Compressed Gas Source or an Anesthetic Vaporizer 1 Connect the Vaporizer Connection Assembly with Bypass to the back of the RoVent Jr 2 Connect the compressed gas or vaporizer to the inlet side of the...

Страница 9: ...a charcoal canister to the exhaust side of the bypass assembly Tubing Configurations Connecting Ventilator Tubing to an Endotracheal Tube Connect the intubation connector tubing directly to the INSP...

Страница 10: ...egin ventilation 3 At the end of the procedure touch Stop ventilator to stop ventilation Allow the animal to recover RoVent Jr calculates the respiration rate and tidal volume using the animal s body...

Страница 11: ...the body weight that determines the automatic settings for respiration rate and tidal volume To change these Parameters manually Touch Respiration rate or Set target volume 2 Choose a ventilator prio...

Страница 12: ...of a breath The cap is on the EXP port The expiratory valve has failed Alarm Messages Out of regulation Over five consecutive breaths peak pressure was more than 25 or less than 75 of your target Requ...

Страница 13: ...ernal leaks as well as leaks in tubing circuits Resulting Message Meaning Possible Action Test OK Tubing leaks No leaks found None is needed The tubing is open to the atmosphere through a loose con ne...

Страница 14: ...s financial obligation extend beyond the cost of or replacement of the Kent product in question Kent Scientific Corporation also reserves the right to make any improvements or modifications to these...

Страница 15: ...RMA number will not be accepted Please note that after our 30 day return policy period ends we will be happy to assist you with your application but cannot issue any credit or refund for a returned R...

Страница 16: ...ntrol Modes Volume controlled pressure controlled Respiratory Rate Range 20bpm to 400bpm Tidal Volume Range 0 01 12 0mL Weight Range 3g to 1 250g RoVent Jr is a registered trademark of Kent Scientific...

Страница 17: ...leak test 10 low baffle pressure 10 main menu 2 4 maintenance 11 more ventilator 2 7 8 param 2 3 parameters 3 4 8 param locations 3 port 9 pressure 2 7 10 13 pressure priority 8 priority menu 2 8 ran...

Страница 18: ...1116 Litchfield Street Torrington Connecticut 06790 Local 860 626 1172 Toll Free 888 572 8887 Part No 10 6000 28...

Отзывы: