Kent Scientific Rovent Jr. Скачать руководство пользователя страница 12

9

RoVent® Troubleshooting

Problem

Possible Cause

Tube open to atmosphere

The highest measured 

pressure in a breath cycle is 

under 25% of the target and 

the tidal volume is over target.

•  Part of the tubing between the INS and EXP ports is open, 

 

disconnected or leaking.

•   There is a large mismatch between the set or calculated tidal 

 

volume and the required tidal volume, or between the set 

 

and actual body weights.

Tube blocked

The highest measured 

pressure in a breath cycle is 

under 25% of the target and 

tidal volume is under target.

•  The endotracheal tube is not seated correctly.

•   There is a kink, blockage or foreign matter in the inspiratory 

 

or expiratory lines.

Exhaust blocked

The breath pressure fails to 

drop during the expiratory 

phase of a breath.

•  The cap is on the EXP port.

•  The expiratory valve has failed.

Alarm Messages

Out of regulation

Over five consecutive 

breaths, peak pressure was 

more than 25% or less than 

75% of your target.

•  Requested Tidal Volume is too high for Pressure and RR .

Large rise in TV

Over the last four breaths, 

there was an increase of more 

than 25% in average tidal 

volume compared to that of 

the previous 16 breaths.

•  Part of the tubing between the INS and EXP ports is open, 

 

disconnected or leaking.

Large drop in TV

Over the last four breaths, 

there was a decrease of more 

than 25% in average tidal 

volume compared to that of 

the previous 16 breaths.

•  There is a kink in the tubing.

•   There is foreign matter in the tubing or in the exhaust valve.

•   The exhaust outlet is blocked.

•   The endotracheal tube is not seated correctly.

Deadspace calib

Dead space calibration 

produced a negative tidal 

volume.

•  Deadspace calibration is incorrect for current tubing setup.

C

F

Содержание Rovent Jr.

Страница 1: ...RoVent Jr Automatic Ventilator Volume Pressure Controlled User s Guide...

Страница 2: ...21 Kent Scientific Corporation Documentation Version 2 00 for On Unit Help Version 4 01 and Firmware Revision 1 01 00 36A April 2021 RoVent Jr Automatic Ventilator Volume Pressure Controlled User s Gu...

Страница 3: ...2 Touchscreen 2 Control 2 Display Options 3 Menus 4 Alarms 4 RoVent Jr Automatic Ventilator 5 Set Up Hardware 5 Ventilating an Animal 7 Configure Settings 8 Troubleshooting 9 General Information 11 Co...

Страница 4: ...orized for use in human applications System Components Front Panel Back Panel Exhaust Air Inlet SERIAL MADE IN U S A 1116 Litchfield Street Torrington CT 06790 Tel 860 626 1172 Toll Free 888 572 8887...

Страница 5: ...ch the bottom half of the Run Screen to open the pop up Display Menu and quickly scroll through charts view other data or to silence Alarms Run Screen Power O General Setup Help Body Weight Start Vent...

Страница 6: ...en Hold down See Main Menu on page 4 Display Options Customizing the Run Screen Display To configure Parameters 1 Touch 1 the upper half of the Run Screen for the Touch Screen Menu 2 General setup 3 C...

Страница 7: ...in Main Menu Screen buttons may vary based on the modules installed Alarms RoVent Jr system alarms alert you to certain conditions that might require attention while RoVent Jr is running There are two...

Страница 8: ...Compressed Gas Source or an Anesthetic Vaporizer 1 Connect the Vaporizer Connection Assembly with Bypass to the back of the RoVent Jr 2 Connect the compressed gas or vaporizer to the inlet side of the...

Страница 9: ...a charcoal canister to the exhaust side of the bypass assembly Tubing Configurations Connecting Ventilator Tubing to an Endotracheal Tube Connect the intubation connector tubing directly to the INSP...

Страница 10: ...egin ventilation 3 At the end of the procedure touch Stop ventilator to stop ventilation Allow the animal to recover RoVent Jr calculates the respiration rate and tidal volume using the animal s body...

Страница 11: ...the body weight that determines the automatic settings for respiration rate and tidal volume To change these Parameters manually Touch Respiration rate or Set target volume 2 Choose a ventilator prio...

Страница 12: ...of a breath The cap is on the EXP port The expiratory valve has failed Alarm Messages Out of regulation Over five consecutive breaths peak pressure was more than 25 or less than 75 of your target Requ...

Страница 13: ...ernal leaks as well as leaks in tubing circuits Resulting Message Meaning Possible Action Test OK Tubing leaks No leaks found None is needed The tubing is open to the atmosphere through a loose con ne...

Страница 14: ...s financial obligation extend beyond the cost of or replacement of the Kent product in question Kent Scientific Corporation also reserves the right to make any improvements or modifications to these...

Страница 15: ...RMA number will not be accepted Please note that after our 30 day return policy period ends we will be happy to assist you with your application but cannot issue any credit or refund for a returned R...

Страница 16: ...ntrol Modes Volume controlled pressure controlled Respiratory Rate Range 20bpm to 400bpm Tidal Volume Range 0 01 12 0mL Weight Range 3g to 1 250g RoVent Jr is a registered trademark of Kent Scientific...

Страница 17: ...leak test 10 low baffle pressure 10 main menu 2 4 maintenance 11 more ventilator 2 7 8 param 2 3 parameters 3 4 8 param locations 3 port 9 pressure 2 7 10 13 pressure priority 8 priority menu 2 8 ran...

Страница 18: ...1116 Litchfield Street Torrington Connecticut 06790 Local 860 626 1172 Toll Free 888 572 8887 Part No 10 6000 28...

Отзывы: