INSTALLATION INSTRUCTIONS

MRIPS312_F-2846_110113

IMPORTANT SAFEGUARDS

When using electrical equipment, basic safety precautions 

should always be followed, including the following:

 

THIS PRODUCT MUST BE INSTALLED IN ACCORDANCE WITH THE 

APPLICABLE INSTALLATION CODE BY A PERSON FAMILIAR WITH 

THE CONSTRUCTION AND OPERATION OF THE PRODUCT AND 

THE HAZARDS INVOLVED. DISCONNECT POWER TO ALL CIRCUITS 

BEFORE WIRING FIXTURE. INSTALL IN ACCORDANCE WITH 

ALL NATIONAL, STATE, AND LOCAL CODES. DO NOT CONNECT 

TO AN UNGROUNDED SUPPLY. READ ALL FIXTURE MARKINGS 

AND LABELS TO ENSURE CORRECT INSTALLATION OF FIXTURE. 

SUPPLEMENTAL INSTRUCTIONS MAY BE LOCATED ON THE 

FIXTURE, IN ADDITION TO THIS INSTRUCTION SHEET, REGARDING 

ORIENTATION, OR MOUNTING RESTRICTIONS. 

 

SAVE THESE INSTRUCTIONS

INSTALLATION INSTRUCTIONS

MEDMASTER

MRIPS SERIES

External Power Supply for MRI Luminaires

This instruction sheet applies for the Kenall MRIPS-312 External Power Supply. It describes the proper mounting 
procedure and wiring instructions for singular and multiple fixture configurations. A separate instruction sheet is 
provided with the MRI luminaire(s). Both sheets together are intended to provide comprehensive instructions on 
system installation and electrical wiring.
The Kenall MRIPS-312 power supply and EMI filter (by others) must be installed outside the shielded MRI 
environment. It contains ferrous components that can damage or interfere with MRI equipment.