INSTALLATION INSTRUCTIONS
MRIPS312_F-2846_110113
IMPORTANT SAFEGUARDS
When using electrical equipment, basic safety precautions
should always be followed, including the following:
THIS PRODUCT MUST BE INSTALLED IN ACCORDANCE WITH THE
APPLICABLE INSTALLATION CODE BY A PERSON FAMILIAR WITH
THE CONSTRUCTION AND OPERATION OF THE PRODUCT AND
THE HAZARDS INVOLVED. DISCONNECT POWER TO ALL CIRCUITS
BEFORE WIRING FIXTURE. INSTALL IN ACCORDANCE WITH
ALL NATIONAL, STATE, AND LOCAL CODES. DO NOT CONNECT
TO AN UNGROUNDED SUPPLY. READ ALL FIXTURE MARKINGS
AND LABELS TO ENSURE CORRECT INSTALLATION OF FIXTURE.
SUPPLEMENTAL INSTRUCTIONS MAY BE LOCATED ON THE
FIXTURE, IN ADDITION TO THIS INSTRUCTION SHEET, REGARDING
ORIENTATION, OR MOUNTING RESTRICTIONS.
SAVE THESE INSTRUCTIONS
INSTALLATION INSTRUCTIONS
MEDMASTER
™
MRIPS SERIES
External Power Supply for MRI Luminaires
This instruction sheet applies for the Kenall MRIPS-312 External Power Supply. It describes the proper mounting
procedure and wiring instructions for singular and multiple fixture configurations. A separate instruction sheet is
provided with the MRI luminaire(s). Both sheets together are intended to provide comprehensive instructions on
system installation and electrical wiring.
The Kenall MRIPS-312 power supply and EMI filter (by others) must be installed outside the shielded MRI
environment. It contains ferrous components that can damage or interfere with MRI equipment.